KANGAROO
e
Pump
™
Enteral Feeding Pump
12. This device is designed to minimize the effects of uncontrolled electromagnetic interference and other types of interference
from external sources. Avoid use of other equipment that may cause erratic operation or degradation in the performance of this
device.
13.
WaRnInG:
The serial I/O port (see Figure 1A) is intended only for use by the manufacturer.
Any accessory equipment that can be connected to the DB9 Serial I/O port, and which does not comply with the equivalent
safety requirements of this equipment, may lead to a reduced level of safety of the pump. While use of the pump’s DB9 port is
not recommended for customers, the following should be considered in the event that the customer decides to connect to this
port:
• The accessory is specifically certified as complying with the safety requirements for accessories used with the pump, whereas
the safety certification of the accessory has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1
harmonized standard.
14.
CaUTIon:
This enteral feeding pump should only be used for patients who can tolerate the flow rates and accuracy levels
delivered by the pump. Premature infants may require higher accuracy rates than specified for this enteral feeding pump.
section III – Icon Identification
Run Mode
Vertically-moving droplet in
RUNNING
screen indicates Run
mode.
battery symbols
• Full charge
• 2/3 or less charge
• 1/3 or less charge
• Very low battery (15 minutes max.)
auto Priming Indicator
When auto priming, the bar will progress
to show auto prime feature is active.
Volume setting
Indicates alarm volume setting.
error Indicator
Pump operations are stopped
until the error is resolved.
Information Indicator
This indicates information only, and
does not require any immediate action.