KANGAROO
e
Pump
™
Enteral Feeding Pump
• Grasp finger tab on valve and insert into left pocket (1).
• Grasp black ring retainer and gently stretch tubing around rotor (2). Pull up on retainer and insert into right pocket (3).
• Press down on finger tab to ensure that valve is fully seated. Finger tab should align with locating rib.
• Close blue door.
(
note:
display status line should read
SET LOADED
).
4. To automatically prime the pump, press
4
PRIMe PUMP
and
4
aUTo PRIMe
. For Feed & Flush sets, auto priming will prime
both lines.
5. To directly control pump priming, press
4
PRIMe PUMP
and then press and hold the
HolD To PRIMe
button(s)
4
. If a Feed
& Flush Set is used, remember to first press
4
HolD To PRIMe flUsH
until the flushing fluid has reached the valve, and then
press
4
HolD To PRIMe feeD
until the feeding fluid has reached past the valve and down to the distal end connector.
6. Use buttons
4
on left side of LCD to set the feeding parameters. Set flushing parameters if a Feed & Flush Set is loaded. The
continuous or intermittent feed mode selection is made under the
MoRe
options menu.
7. Press
4
RUn
when ready to start. The screen will display
RUNNING
.
8. To stop, press
4
HolD
or press and hold the
PoWeR
button.
figure .
Installation of
Pump set.
black Ring
Retainer
(MIsTIC)
finger
Tab
Valve
Rotor
General startup
Placement/Mounting
There are two recommended placement methods for the KANGAROO ePump enteral feeding pump.
• Attached to a vertical IV pole via the Pole Clamp peripheral device, included with the pump.
• Placed feet down on any horizontal surface.
a/C Power operation
Plug the KANGAROO ePump enteral feeding pump into an A/C power outlet for normal operation. When the pump is not plugged in,
or if A/C power is interrupted, a built-in rechargeable battery will run the pump.