T6075 Online Suspended Solids Analyzer
Operating Manual
Shanghai Chunye Instrument Technology Co.,Ltd
25.0
°C
220
mV
4.01
mg/L
2019-01-09 12:53:17
Suspended Solids
T6075
Page 1: ...ne Suspended Solids Analyzer Operating Manual Shanghai Chunye Instrument Technology Co Ltd 25 0 C 220 mV 4 01mg L 2019 01 09 12 53 17 Suspended Solids 25 0 C 220 mV mg L 2019 01 09 12 53 17 Suspended Solids T6075 ...
Page 2: ...Installation Installation size Instrument installation Instrument connection Keypad descriptions Display descriptions Menu structure Calibration Graphic Trend MODBUS RTU Daily Maintenance Package Set FAQ Warranty Notes 1 1 2 3 3 4 5 6 7 8 10 12 13 20 20 21 22 23 ...
Page 3: ... the instrument and accessories are damaged by transportation and whether the accessories are complete If any abnormalities are found please contact our after sales service department or regional customer service center and keep the package for return processing This instrument is an analytical measurement and control instrument with highly precision Only skilled trained or authorized person shoul...
Page 4: ...perature Relative humidity Waterproof rating Weight Dimensions Installation opening size Installation methods 0 500 5000mg L 0 50 100g L Can be extended mg L g L 0 001mg L 0 1g L 1 F S 0 50 0 1 0 3 Two 4 20mA 20 4mA 0 20mA RS485 MODBUS RTU Data record Curve display 5A 250VAC 5A 30VDC 85 265VAC 9 36VDC power consumption 3W No strong magnetic field interference around except the geomagnetic field 10...
Page 5: ...Instrument installation Instrument dimensions Back fixed hole size Embedded mounting Cut out size 138 mm 138 mm 0 5 mm 0 5 mm M4x4 45x45 mm 144 mm 136 mm 144 mm 26 mm 118 mm 3 Installation size ...
Page 6: ...n diagram a Embedded in the opened hole b Install a fixed frame for the instrument a Install fixing bracket for instrument b Wall screw fixing Instrument installation Wall mounted installation D D B2 L N RELAY A RELAY B RELAY C Install fixed bracket top view pay attention to 4 6x15 mm 150 3 mm 58 1 mm ...
Page 7: ...eters insert the corresponding label or color wire on the sensor into the corresponding terminal inside the instrument and tighten it V V A1 B1 V V A2 B2 I1 G I2 A3 B3 G TX RX P P RLY3 RLY2 RLY1 L N Digital input channel 1 Digital input channel 2 Output current RS485 communication output RS232 communication output data upload DC power supply 3 sets of relays L FireWire N Neutral Ground V V A1 B1 V...
Page 8: ...rt press means to release the key immediately after pressing Short press if not specified below Long press Press and hold for 3 seconds and then release the key Menu setting mode press this key to loop down the menu options Calibrated check the calibration status Re calibration press ENT again Menu setting mode press this key to rotate menu options Press the key in measurement mode to display the ...
Page 9: ...the power is switched on the meter will display as follows Relay 1 blue is off and red is on Relay 2 blue is off and red is on Relay 3 blue is off and red is on Current 1 Current 2 Switch display Automatic temp comp Temp Unit Voltage value corresponding to the main measured value 2019 01 09 12 53 17 Suspended Solids 14 51 mA C1 SP1 SP2 SP3 5 mV Voltage Calibrating Point 1 mg L 0 3 25 0 C 4 01mg L ...
Page 10: ...ual C F 0 01 Default can be modified 1 Default can be modified 10 Default can be modified 100 Default can be modified 1000 Default can be modified Voltage 1 Voltage 2 Voltage 3A Voltage 3B Voltage 4A Voltage 4B Voltage 5 ON OFF High Alarm Low Alarm Cleanning Cleaning time setting is as below Continuous opening time The interval between the last opening and closing and the next opening ON OFF High ...
Page 11: ... Data Query Interval Memory information Data Output Language Date Time Display Soft Version Factory Default Terminal Current Tuning Relay Test Output Data Log System 9 Main Temperature 4 20mA 0 20mA 20 4mA Main Temperature 4 20mA 0 20mA 20 4mA 4800BPS 9600BPS 19200BPS None Odd Even 1 Bit 2 Bit 001 Low Standard Medium High Saving Bright First gear Second gear Third gear Automatic 19 1 0 0000 The po...
Page 12: ...Standard solution calibration has five points and can be calibrated at any point at least one point In the standard solution calibration mode press key to switch correction points and press ENT key to start correction If the instrument has been calibrated press the CAL key to check the calibration state press the key to switch the calibration state of the calibration point and if the point shall b...
Page 13: ...d Calibration Calibrating mg L 07 00 Vol 002 Temp 25 0 C Slope 00 984 Calibration results Confirm When the ENT icon is green press ENT to confirm Cancel Press the Ӝ key to shift the green icon to ESC and press ENT key to confirm Offset adjustment Compare the data of laboratory or portable instrument with the data of instrument measurement if there are errors the error data can be modified by this ...
Page 14: ...ction of displacement Interval point 1h point 12h point 24h point year month day 7 5s 90s 180s Curve query trend chart Data Query Interval Data Log Graphic Trend Trend Chart Press սMenuվ key to enter the setting mode set the recording interval and the instrument will 400 points per screen displays the most recent data trend graph according to interval Settings 400 points per screen display trend c...
Page 15: ...introduces additional fields of protocol data units The client that initiates the MODBUS exchange creates the MODBUS PDU and then adds the domain to establish the correct communication PDU Figure 2 MODBUS architecture for serial communication On the MODBUS serial line the address domain contains only the slave instrument address Tips The device address range is 1 247 Set the device address of the ...
Page 16: ...TU mode message frames are distinguished by idle intervals of at least 3 5 character times which are called t3 5 in subsequent sections Figure 5 RTU message frame The entire message frame must be sent in a continuous character stream When the pause time interval between two characters exceeds 1 5 characters the information frame is considered incomplete and the receiver does not receive the inform...
Page 17: ...ad and hold register request frame Response N Register number Figure 9 Read and hold register response frame The following illustrates the request frame and response frame with the read and hold register 108 110 as an example The contents of register 108 are read only with two byte values of 0X022B and the contents of register 109 110 are 0X0000 and 0X0064 Value Byte Start No signal bytes during 3...
Page 18: ...ction code Start Address Number of input registers number of bytes Register values 1 byte 2 byte 2 byte 1 byte N 2 byte 0x10 2 byte 2 byte 1 byte N 2 byte Request Frame Number Systems Function code Start address high byte Start address low byte Input register number high bytes Input register number low bytes number of bytes Register value high byte Register value low byte Register value high byte ...
Page 19: ...ly the decimal point behind can be recorded For the important explanation of 23 bit mantissa the first i e hidden bit is not compiled Hidden bits are bits on the left side of the separator which are usually set to 1 and suppressed Step 4 Symbol bit definition The sign bit of positive number is 0 and the sign bit of negative number is 1 so the sign bit of 17 625 is 0 Step 5 Convert to floating poin...
Page 20: ...0 0 4 0 0 132 Mantissa bits M 11110110110011001100110B 8087142 Step 2 Calculate the decimal number D 1 𝑆 1 0 M 223 2𝐸 127 1 0 1 0 8087142 223 2132 127 1 1 964062452316284 32 62 85 Reference Code float floatTOdecimal long int byte0 long int byte1 long int byte2 long int byte3 long int realbyte0 realbyte1 realbyte2 realbyte3 char S long int E M float D realbyte0 byte3 realbyte1 byte2 realbyte2 byte1...
Page 21: ...ter addresses Communication address is as follows 0000 0001 Temperature value 0002 0003 Main Measured Value 0004 0005 Temperature and Voltage Value 0006 0007 Main Voltage Value Communication examples Examples of function code 04 instructions Communication address 1 temperature 20 0 Main value 10 0 temperature voltage 100 0 Main voltage 200 0 Host Send 01 04 00 00 08 F1 CC Slave Response 01 04 10 0...
Page 22: ...res otherwise the instrument may be damaged or its service life may be reduced 3 When cleaning the plastic shell of the instrument please use a soft cloth and a soft cleaner to clean the shell Be careful not to let moisture enter the inside of the instrument 4 Check whether display data of the instrument is normal or not 5 Check whether the wiring on the terminal of the instrument is firm Pay atte...
Page 23: ...calibrated Solutions Please re calibrate the 20mA output 4 The instrument shows Possible causes The transmitter and sensor are not communicating properly Solutions Check that the sensor signal cable is correct Please refer to the wiring diagram in the manual 5 Measurement shows the result as full scale SLOP Possible causes May be sensor was contaminated damaged or measured value exceeds measuremen...
Page 24: ... reason When applying for authorization please nclude date requiring the reason of return Instruments must be carefully packed to prevent damage in shipment and insured against possible damage or loss We Instruments will not be responsible for any damage resulting from careless or insufficient packing Warning Damage as a result of inadequate packaging is the User distributor s responsibility Pleas...
Page 25: ...se it should be stored in a cool dry and ventilated place ӹ If you don t use the instrument for a long time you should unplug the power supply to avoid accidents ӹ This instrument is not suitable for use in harsh environment high temperature low temperature or strong magnetic field interference which may lead to instrument damage ӹ If there is any problem with the instrument please contact the dea...
Page 26: ...Shanghai Chunye Instrument Technology Co Ltd Tel 021 61621082 Fax 021 61621099 Web www twinno net Add Building 6 No 166 Mindong Road Pudong New District Shanghai China Post code 201209 ...