Page 48
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
Forced cooling to the required end temperature, chamber
pressure remains at approx. 300 kPa and the cooling valve is opened.
During the interval of the cooling stage, between 121°C and
100°C the cooling rate is kept to approximately a decrease of 1
degree per minute. This will make the last stage of the cooling
take approximately 21 minutes. This is to check the durability of
the bottles.
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.15
Program 15: Durham
For liquids when the manufacturer recommends autoclaving at temperatures
of 121ºC (250
F) for 15 minutes.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature: 121
C (250
F).