Page 63
2.
0000 is displayed on the screen with the cursor blinking on the right
digit.
3.
To increase or decrease the digits, press the
UP
or
DOWN
keys.
4.
Set your password, then move the cursor to
Set
by pressing the
START/STOP
key.
5.
When
Set
is blinking, press the
UP
or
DOWN
keys to return to the
program screen.
The autoclave starts performing sequence of operations. The actual
measured values of pressure and temperature are displayed
continuously and printed every minute at STE stage, and every 5
minutes at the other stages. The phase in progress is displayed at the
right side of the upper line as WATER, HEAT, STER., and EXH.
If the operator presses the START key and the door is not completely
closed, the process will not start and the DOOR indicator will flash
twice then turn off and the buzzer will sound four times.
Caution!
Do not touch the strainer’s cover, mounted on the exhaust line,
during and short after operation.
Touching the hot strainer’s cover may cause severe injuries.
10.5
Unloading
•
When the cycle ended successfully (including pressing the
START/STOP
key, or any failure, after completing the sterilization
stage) message "Cycle Ended" (and the relevant failure message, if
applicable) is displayed on the screen.
•
Verify than there is no pressure in the chamber, according to the
reading on the display. Only then you may open the door.
Warning
To avoid severe injuries from hot steam when opening the door:
It is strictly forbidden to lean on the autoclave.
It is strictly forbidden to place your hand or any part of your
body over the door.