Page 62
▪
Compatible material
▪
Proper weight.
10.4
Operations
1.
Select the program.
▪
UP
key: next program.
▪
DOWN
key: previous program.
Attention:
Selecting a program is possible only when the door is open.
2.
Verify that you chose the required cycle.
3.
If the autoclave is equipped with a printer verify that a paper roll is
inserted in the printer. If not - insert as per para. 5.2.
10.4.1 Closing the door
1.
Hold the opening handle in open position, while pushing the door until
it comes to closed position, then release the handle.
2.
The open door symbol
is replaced by the message "System
Ready".
10.4.2 Starting cycle
Start the cycle by pressing the START/STOP key.
If your autoclave supports CFR 21 part 11 standard, perform the following
procedure:
SELECT USER
screen will be displayed:
1.
Enter the Enter Code screen by moving the cursor to your username
(Use
Up
and
Down
keys) and pressing
START/STOP
key. The
following screen will be displayed: