Page 87
add water to the reservoir until the message disappears. Do not fill the
reservoir completely!
6.
Put a few drops of oil on the two door pins and door tightening bolts.
1.
Clean outer parts of the autoclave with a soft cloth.
11.1.3
Periodically
1.
Every 6 months replace the air filter, (if installed) (to be done by an
authorized technician).
2.
Once a year check and tighten the piping joints to avoid leakage (to be
done by an authorized technician).
3.
Once a year check and tighten all screw connections in the control box,
heaters, valves, and instrumentation (to be done by an authorized
technician).
4.
Check the door gasket every 12 months and replace it if required (refer
to the technician manual). Replacing the gasket shall be done by a
technician.
11.1.4
Periodical Tests
1.
Once every month activate the safety valve (see sec. 11.2).
11.2
Checking the Safety Valve
The safety valve is located on the rear side of the autoclave
To prevent the safety valve from a blockage, operate it once a month.
11.2.1
PED-approved type safety valve
Operate the sterilization cycle according to the manual.
Allow a pressure of approximately 200 kPa (29 psi) to build up in the
chamber.
Turn the safety valve pressure-regulating nut clockwise for 2 seconds.
Be careful not to burn your hands.
Return the nut to its original position.
Press the
STOP/STOP
key to interrupt the operation, and exhaust
steam from chamber.
Wait until the pressure decreases to zero, only then can the door be
opened.
Attention:
Summary of Contents for LabSci 11-V
Page 19: ...Page 18 8 Right service door lock...
Page 22: ...Page 21 10 Drain outlet...
Page 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Page 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Page 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Page 79: ...Page 78...