Page 99
10.8.
Cleaning water strainer
CAUTION!
Do not touch the strainer or chamber shortly after operation as
they will be very hot.
Touching the hot strainer may cause severe injuries. If this
maintenance operation is performed while the strainer is hot, use
heat resistant gloves to avoid injuries.
1.
Turn the unit on momentarily to allow the door to be opened. Then turn
the unit off to stop the chamber from preheating.
2.
Remove the trays and the tray holder.
3.
Remove the strainer element mounted on the bottom-rear of the
chamber.
4.
Clean the strainer with water using a nylon or brass bristle brush if
necessary.
5.
Reinstall the strainer element.
6.
Reinstall the tray holder and the tray.
The strainer element is located
on the back of the chamber
Strainer element