8
2.6
Utilities
Utilities
Unit
Value
Power supply (as appropriate)
V-A 1ph, 120V – 16A,50/60 Hz
V-A 1ph, 230V – 16A,50/60 Hz
Attention:
1.
The electrical net must be protected with a current leakage safety
relay (GFI Receptical or Circuit Breaker).
2.
The electrical network must comply with local rules or regulations.
3.
The autoclave must be connected to a properly grounded outlet.
2.7
Waste Water Disposal
Caution !
Waste water must be brought into the public water piping in
accordance with the local rules or requirements, i.e. only non-
hazardous liquids may be disposed in public sewage!
2.8
Environment Emission Information
A. The peak sound level generated by the sterilizer is
<
70 / dBA with
background noise of 60 dB.
B. The total heat transmitted by the sterilizer is < 100 W/h for
1730/2340/2540 models and < 150 W/h for 3140/3850/3870
models.