6
2.2
Storage Conditions
The packed or unpacked autoclave shall be stored in “indoor
conditions” (protected from rain and water).
2.3
Operating Conditions
This device is intended for indoor use.
This autoclave is intended for NORMAL environment conditions as
follows:
Altitude up to 2000m.
Minimum room temperature 41ºF (5ºC).
Installation Category II.
Pollution Degree 2.
Maximum relative humidity 80% for temperature up to 31
º
C
decreasing linearly to 50% relative humidity at 40
º
C.
Mains supply voltage fluctuations up to ±10% of nominal voltage.
The sterilizer should be loaded only with autoclavable material.
2.4
Construction
The main parts of the autoclave are made of materials as indicated
below:
Chamber is built of electro-polished stainless steel 316 L.
Door is made of stainless steel CF8.
Trays are made of stainless steel 304.
Door handle is made of hard plastic material that is safe to touch
and thermo-insulated.
Water reservoir is made of hard plastic material.