13
Function and Programming
No
Parameter
Description
Range
Default
01
Lock
Software Lock
On / Off
On
02
Address
Address on MODbus
Off, 1…247
Per Order
03
Baudrate
MODbus baudrate
9,600 / 19,200
Per Order
04
Analogue Output Type Toggle for Current Loop Type
4-20mA, 0-20mA
Per Order
05
Analogue Output
Mode
Select Polarity of Analogue Output
noninverted,
inverted
Per Order
06
0/4 mA Whole
Scale Low setpoint for output – Whole
Percent
0% to 98%
Per Order
07
0/4 mA Dec.
Scale Low setpoint for output –
Decimal Point 0-97.XX
XX.00% - XX.99%
Per Order
08
20 mA Set
Scale High setpoint for
output – Whole Percent
2% to 100%
Per Order
09
20 mA Set
Scale High setpoint for output – Decimal
Point 2-99.XX
XX.00% - XX.99%
Per Order
10
D.O. Units Selected
Select between ppm and % Saturation
units for output
ppm or % Sat
Per Order
11
Conductivity Units
Selected for Output
If Conductivity Sensor Type = 6 (Standard/
High) then choices are uS/cm, Salinity
(PSU) or TDS If Conductivity Sensor Type
= 7 (Ultralow) then choices are uS/cm, MΩ
Standard or MΩ for UPW
For Sensor Type =
6 uS/cm, PSU, TDS
or For Sensor Type
= 7 uS/cm, MΩ,
UPW
Per Order
12
Conductivity Sensor
Cell Constant (K)
Indicates nominal cell constant for connected
sensor From K=0.01/cm to =20.00/cm
0.01 to 20.0
Per EC Sensor
13
Conductivity Sensor
Range Mode
Indicates the range mode scaling factor for
EC sensor “UL”=2, “Std”=200, “Hi”=2,000
2..2,000
Per EC Sensor
14
0/4mA Offset
Trim Low
±9.99% *
Per Factory Cal
15
20mA Gain (Span)
Trim High
±9.99% *
Per Factory Cal
16
Sampling Rate
Set sampling frequency in seconds
0.5, 1.0, 2.0 & 4.0 Per Order
17
Recalibrate Notify
Set max time since cal last peformed be
-
fore notification
1 to 999 Days
Per Order
18
Display Sensor Type
1=pH, 2=ORP, 3=Wide ORP, 4=DO,
5=ISE, 6=Cond Standard/High Style,
7=Cond Ultralow Style
1..7
Per Sensor
19
Formula Weight
Formula Weight of Measured Ion –
Only for ISE sensors
6.94..655.35
Per ISE
Sensor
20
Type of TDS Units
Type of TDS units which are sent from EC
Sensor
0=NaCl,
1=442,
2=KCl
Per EC Sensor
21
Back to Default
Reset to Default
Def = Reset,
Par = No Reset
Par
*Negative values will be shown as flashing. Shaded portions of chart above indicate display only parameters.