APPENDIX D: TROUBLESHOOTING
TD1000C™Oil in Water Monitor
32
TD-1000C Oil In Water Monitor
Operation Manual
PN 101687
REV G
SYMPTOM
POSSIBLE CAUSE
SOLUTION
HOME
screen reads zero.
System Values incorrectly set
Make sure valid System Values
were entered.
Screen blank or black.
LCD contrast too high or too low
If screen is blank, adjust contrast
by pressing UP ARROW (if screen
is black, use the DOWN ARROW)
continuously until screen is visible.
Use ñò arrows to fine adjust.
Monitor does not calibrate.
Failure to complete entire calibration
procedure
You must press <1> at the end
of the calibration sequence for
the monitor to accept the values.
Recalibrate.
AC Relay is not operating prop-
erly.
High Signal Alarm setting is incor-
rect or terminal strip connections are
intermittent.
If there is an alarm, check the high
signal level on monitor. Refer to
Diagnostics section; relay test func
-
tion, to test if the monitor is properly
activating the control device.
Fuse has “blown.”
Replace fuse with spare located on
the Main PCB.
Offset value ineffective (i.e.
HOME
screen reading does not
change when the Offset value is
changed).
This is not usually cause for alarm
as the
HOME
screen reading is a
result of a combination of factors.
Do not attempt to change current
settings unless you are certain
something is wrong. Consult
Turner Designs Hydrocarbon In-
struments at +1 559 253 1414.
Monitor does not respond to sec-
ondary reference solutions.
System Values incorrectly entered
Insure correct Cal Solution, Offset
and Fluor Temp Comp. Value have
been entered.
Cell calibration alarm
Dirty flow cell
Clean flow cell. Re-calibrate cell
condition.
D.4 TD1000C™ TROUBLESHOOTING WORKSHEET
The Troubleshooting Worksheet is designed to facilitate diagnostic data collection. After collecting
all data on the worksheet, most problems can be solved over the telephone. Refer to the Service
Assistance Section for the appropriate numbers.
Please fill out worksheet completely -- All entries are important