CALIBRATION
TD1000C™Oil in Water Monitor
20
TD-1000C Oil In Water Monitor
Operation Manual
PN 101687
REV G
The entire calibration procedure must be completed to store the new values entered during
calibration.
5.4 CALIBRATION DATA SCREEN
This screen provides a check on proper fluorescence calibration. It is accessed from the
HOME
screen
by pressing <
ENT
>, then <
2
>. See
Table 1
for default values and ranges.
Blank: Shows raw data output for blank solution as set during calibration. It can be used to check
proper calibration.
Cal Std: Shows raw data output for the standard solution as set during calibration. It can be used to
check proper calibration.
Table 1—Calibration Data
Parameter
Order of Key Presses
Default
Range
Blank
<
ENT
><
2
>
0.0
0–250
Cal Std
<
ENT
><
2
>
500.0
0–1000
5.5
FLUORESCENCE CALIBRATION SCREENS
HOME
XXXX RFU T:
23C Cell: 100%
1. From the
HOME
Screen press <
ENT
>.
ENT
1. Calib 2. Data
Cal XX Days Ago
2. Press <
2
> to view raw data output for blank and cal standard as set during
current calibration, OR to begin calibration sequence.
2
Blank: XX.X Cal
Std. XXX.X
1
Please Input ID:
3. If requested, key in ID and press <
ENT
>. If not requested, go to step #4.
ENT
CALIBRATION SYSTEM Press
<1> to Start
4. Press <
1
> to begin.
1
BLANK SOLU-
TION
5. Close the inlet valve to stop flow by moving handle to horizontal position.
Inject 120 mL of blank solution (flush with 60 ml, and then allow portion of
second 60 ml to remain in flow cell). Press <
ENT
>
ENT
BLANK: XX
6. If the number shown is less than 25, wait for the number to stabilize, the
press <
0
>.
Ý
BLANK: XX
wait/WAIT
7. After
WAIT
command disappears, monitor will prompt you to press <
ENT
>
to continue.
To Continue Press
<ENT>
8. Press <
ENT
>.