EO-25X

Instruction

Manual

Oil/Liquid Temperature Calibrator

Temperature Range: -25˚C to 100˚C

Tunix Corporation, MS1-A, Opp mall godown road, New siyaganj, Indore 452001, MP, India

Summary of Contents for EO-25X

Page 1: ...unix Corporation EO 25X Instruction Manual Oil Liquid Temperature Calibrator Temperature Range 25 C to 100 C Tunix Corporation MS1 A Opp mall godown road New siyaganj Indore 452001 MP India...

Page 2: ...Table of Contents 1 Preface 1 2 Technical Specification 2 3 Operating Principle 3 4 Safety Measures 5 5 Trouble Shooting 7...

Page 3: ...ive you trouble free performance for longer period Before starting the instrument we suggest you to go through the instruction manual Please check for following Standard accessories 1 Product is not p...

Page 4: ...al Non uniformity Axial 0 09 C after stabilization time of 15 minutes 7 Stabilization Time settling time 15 minutes after set point is achieved 8 Time to reach to 100 C from ambient temperature 30 min...

Page 5: ...he calibration depends on 1 Stability of the source 2 Uniformity to which the stable temperature is known Designed block diagram of EO 25X is given below 230VAC 50 Hz PID Controller Control Output SSR...

Page 6: ...E 1 Use 10 Amp glass fuse as supplied with instrument 2 3 Pin Plug used should have 15 amp capacity 3 Cooling fan at the backside of controlling unit should be always on PID Controller This is require...

Page 7: ...in higher non uniformity Always use correct platform to keep the instrument Incorrect level may increase the vibrations Never carry transport the product with filled oil vessel Never carry transport...

Page 8: ...re of heating element In case of fuse being fused to frequently kindly consult factory In the event of SSR failure temperature gets run away above set value If it cross the set point by more than 100...

Page 9: ...it Check for correct polarity in your mains plug Phase should go through SSR as per design If phase is not routed through SSR It may not control the temperature 3 Chamber is not not getting on Check m...