Function
Parameter
Settings/Options
Weighing Unit
Decimal Setting
Maximum Capacity
Zero Calibration
Graduation Setting
(readability of the least
significant digit)
C01
C02
C03
C04
C05
1 = kg
2 = lb
1 = kg
2 = lb
3 = gram
4 = oz
0 = no decimal
1 = 0.0
2 = 0.00
3 = 0.000
4 = 0.0000
options: 1/2/5/10/20/50
Example with no decimal places (ie. C02=0)
1 = 1 lb
2 = 2 lb
5 = 5 lb
10 = 10 lb
20 = 20 lb
50 = 50 lb
set max capacity
0 = zero calibration not needed
1 = set the zero calibration (Please ensure scale is empty and
the stable light is on)
Note: for calibration only kg or lb are allowed
Model 1
Model 2
INDICATOR PARAMETER SETTINGS
The parameter settings menu has a calibration section (C01 to C07 explained above) and a
parameter settings section (C08 and up).
To access the calibration section the seal switch (located at one corner of the PCB) must be
OFF. This will allow access to all C01 and up settings. If the seal switch is ON, then only C08
and up can be accessed by the user. If you break the official seal by opening the back of the
indicator to access the seal switch, you may need to have the indicator recertified. Be sure to
adjust the seal switch back to the original setting after calibration/configuration has been
performed.
To enter calibration/parameter settings, follow the procedure below:
1. Make sure UNITS is set to either kg or lb
2. Press and hold
HOLD
and
key at the same time for 2 seconds
3. Navigate through the settings (C01 to C45) as shown in the table 4 below by using the arrow
keys and return keys as labeled under each indicator button
4. Press
key to enter/edit the parameter setting
5. Press
PRINT enter / HOLD esc
key to save and exit settings at any time
- 9 -
Calibration
C06
0 = calibration not needed
1 = Ready to calibrate with one calibration weight
2 = Ready to calibrate using multiple calibration weights (Linear)
3 = Sensitivity Output