Features:

• Wavelength Response 950 nm to 1650 nm
• Low Noise, High Gain
• Gain Settings from 2 to 7
• Bandwidth 30 KHz to 7 GHz typical
• SMA type K Output Electrical Connector  
• FC/APC Style Fiber Optic Input Connector

Made In the USA

©10/2014 Terahertz Technologies Inc.

Optical to Electrical Converter

T

I

A

4
0
0
0

T

he TIA-4000 is comprised of a 

fi

 ber coupled InGaAs APD detector combined with a variable reverse bias 

voltage network and fast transimpedance ampli

fi

 er. The output of the unit brought out to a type K female 

SMA connector. Light falling on the detector generates a positive-going proportional current. This current, 
multiplied by the transimpedance produces a voltage that is proportional to the light incident on the detector 
surface.

Normally the unit is used to drive a coaxial cable, this cable should have a 50 ohm characteristic impedance 
and be terminated with a 50 ohm load at the oscilloscope or other measuring device to be used. 

The active area of the TIA-4000 is 30 microns in diameter. It is coupled to a single mode optical 

fi

 ber. Typical 

optical return loss is 52 dB and is caused primarily by the FC/APC connector employed. 

The applied bias voltage controls the gain of the APD. This voltage in turn, is controlled by means of the two 
push buttons on the top panel. The rightmost button increases the gain and the leftmost button decreases the 
gain. There are 64 discrete gain settings that are retained in non-volatile memory when the power is turned off. 
Thus the gain setting will be retained even though power is disconnected from the unit. 

Pressing a button once causes the unit to advance in the selected direction by one step. Holding the button 
down causes the unit to advance automatically until a limit is reached.

Operation Notes:
1:   To avoid damaging the detector, the gain must be set to the lowest setting prior to turning off the TIA-4000.

2:   Do not apply optical power to the detector until AFTER the device is power on and the APD bais is con

fi

 rmed                             

to be set to the lowest level. 

TIA Freq. Response

GHz

1

 

2

 

3

 

4

 

5

 

6

 

7

 

8

 

9

0

-2

-4

-6

-8

-10

Re

la

tive Re

sp

o

n

se

 -

d

B

10

Terahertz Technologies Inc.

169 Clear Rd, Oriskany  NY  13424

855-TTI-TEAM (884-8326)

Phone: 315-736-3642 Fax: 315-736-4078

email: [email protected]

web: www.teratec.us