General 

Initial Operation 

This section is a general introduction to the organisation of the instrument and is intended to be 
read before using the generator for the first time. Detailed operation is covered in later sections 
starting with Standard Waveform Operation. 

In this manual front panel keys and sockets are shown in capitals, e.g. STATUS, AUX OUT; all 
soft

key labels, entry fields and messages displayed on the LCD are shown in a different 

type

font, e.g

.  

WAVEFORM

,  

sine

DDS Principles 

In this instrument waveforms are generated by Direct Digital Synthesis (DDS). A  phase 
accumulator is incremented at a rate proportional to the required output frequency. The most 
significant 12 bits of the accumulator are used to address a look-up table ROM that converts the 
phase information into sinewave amplitude data; this data is then passed to a 10-bit Digital-to-
Analogue Converter (DAC) which produces the output waveform. For triangle waves the ROM 
look-up table is by-passed and the phase accumulator output is passed directly to the DAC. 

At low frequencies all 4096 points in the output wave are stepped through, but as the frequency 
increases points are progressively missed out. Sinewaves and triangles are subsequently filtered 
to smooth the steps in the DAC output; this technique ensures good sinewave purity up to the 
maximum frequency of the generator but the practical limit to which excellent triangle linearity is 
maintained is only about 100kHz. Squarewaves and pulses are derived from the sinewave using 
a variable threshold comparator; this permits symmetry control of these waveforms across the 
whole instrument frequency range. 

The major advantages of DDS over conventional analogue generation are: 

• 

Frequency accuracy and stability is that of  the crystal oscillator. 

• 

Frequencies can be set with high resolution from mHz to MHz. 

• 

Low phase noise and distortion. 

• 

Very wide frequency sweeps are possible. 

• 

Fast phase continuous frequency switching. 

• 

In addition, being a digital technique, it is easier to make every parameter programmable from 
the keyboard, or remotely via the USB or RS-232 interfaces. 

Switching On 

Switch on the generator using the ON/OFF switch on the rear panel. To fully disconnect from the 
AC supply unplug the mains cord from the back of the instrument or switch off at the AC supply 
outlet; make sure that the means of disconnection is readily accessible. Disconnect from the AC 
supply when not in use. 

At power up the generator displays the installed software revision and conducts a self-test. 
Power-on self-test takes a few seconds, after which the status screen is displayed, showing the 
generator parameters set to their default values, with the MAIN OUT output set off. Refer to the 
System Operations section for how to change the power up settings to either those at power 
down or to any one of the stored settings. Recall the status screen at any time with the STATUS 
key; a second press returns the display to the previous screen. 

Change the basic generator parameters as described in the Standard Waveform Operation 
section and switch the output on with the ON key; the lamp above the key will light to show that 
output is on. 

 12 

Summary of Contents for tg2000

Page 1: ...THURLBY THANDAR INSTRUMENTS 20MHz DDS FUNCTION GENERATOR TG2000 INSTRUCTION MANUAL...

Page 2: ...utput 18 Warnings and Error Messages 19 Sweep Operation 20 General 20 Setting Sweep Parameters 20 Gated Mode 25 Tone Mode 26 FSK 27 Modulation 28 System Operations from the Utility Menu 29 Calibration...

Page 3: ...us Sweep can be linear or logarithmic single or continuous Single sweeps can be triggered from the front panel the trigger input or the digital interfaces A sweep marker is provided AM External Amplit...

Page 4: ...digits Symmetry Control 20 to 80 1 resolution 1mHz to 20MHz Accuracy 10 ppm for 1 year 1mHz below 0 2Hz Output Level 2 5mV to 10Vp p into 50 Rise and Fall Times 22ns Aberrations 5 2mV Triangle Range 1...

Page 5: ...VCA IN socket Frequency Shift Keying FSK Phase coherent switching between two selected frequencies at a rate defined by the switching signal source Carrier frequency From 1Hz to 20MHz Carrier wavefor...

Page 6: ...100kHz for Gate DC 2 5kHz for Tone and Sweep Signal Range Threshold nominally TTL level maximum input 10V Minimum Pulse Width 100ns for Gate and FSK modes 0 2ms for Sweep and Tone modes Input Impedan...

Page 7: ...interruption is prohibited The protective action must not be negated by the use of an extension cord without a protective conductor When the instrument is connected to its supply terminals may be liv...

Page 8: ...ropean Communities EMC Emissions a EN61326 1 2006 Radiated Class B b EN61326 1 2006 Conducted Class B c EN61326 1 2006 Harmonics referring to EN61000 3 2 2006 Immunity EN61326 1 2006 Immunity Table 1...

Page 9: ...2004 Fast Transient 1kV peak AC line 0 5kV peak signal connections Performance A B e EN61000 4 5 2006 Surge 0 5kV line to line 1kV line to ground Performance A B f EN61000 4 6 2007 Conducted RF 3V 80...

Page 10: ...only 5 Reassemble in the reverse order 6 To comply with safety standard requirements the operating voltage marked on the rear panel must be changed to clearly show the new voltage setting 7 Change the...

Page 11: ...the falling edge is at the 180 phase point For square waves and pulses both phase and symmetry are coincident with MAIN OUT Trigger Provides a replica version of the actual trigger signal internal ext...

Page 12: ...ta 8 TXD2 Secondary transmitted data 9 GND Signal ground Pin 2 3 and 5 may be used as a conventional RS 232 interface with XON XOFF handshaking Pins 7 8 and 9 are additionally used when the instrument...

Page 13: ...re derived from the sinewave using a variable threshold comparator this permits symmetry control of these waveforms across the whole instrument frequency range The major advantages of DDS over convent...

Page 14: ...reen a new frequency of 100kHz for example is set by pressing 1 0 0 kHz CE Clear Entry undoes a numeric entry digit by digit ESCAPE returns a setting being edited to its last value Eight soft keys aro...

Page 15: ...hanged by further presses of the soft key by pressing either cursor key or by using the rotary control For example pressing the AUX OUT soft key on the main menu brings up the screen shown below AUX O...

Page 16: ...is selected the output waveform is the baseline DC voltage only set from the DC Offset screen Frequency FREQUENCY 10 0000 kHz freq period Pressing the FREQ key gives the FREQUENCY screen With freq se...

Page 17: ...he rotary control will increment or decrement the numeric value in steps determined by the position of the edit cursor flashing underline the cursor is moved with the left and right arrowed cursor key...

Page 18: ...attenuation provided by the fixed attenuator using the values determined during the calibration procedure In the example displayed the output signal is 250mVpp exactly and takes account of the small e...

Page 19: ...nt sections later in this manual Pressing the AUX OUT key calls the AUX OUT setup screen AUX OUT output on mode auto srce waveform sync AUX OUT is turned on and off by alternate presses of the output...

Page 20: ...shown when an illegal setting is attempted most generally a number outside the range of values permitted In this case the entry is rejected and the parameter setting is left unchanged Examples are 1 E...

Page 21: ...off either by the sweep soft key on the MODE screen accessed from the Main menu by the on and off soft key on the SWEEP SETUP screen accessed by pressing setup on the MODE screen or by the on and off...

Page 22: ...quencies can be entered with a resolution of 0 1Hz or 5 digits but in sweep mode the instrument operates with an increment resolution of 0 2Hz and the actual frequency of any particular step will be r...

Page 23: ...sync set to off the frequency steps directly and phase continuously from the stop frequency to the start frequency but is not synchronised to the software generated trigger signal In triggered mode t...

Page 24: ...rotary control Sweep Spacing Pressing the spacing soft key on the SWEEP SETUP screen calls the SWEEP SPACING screen SWEEP SPACING off logarithmic linear done With linear selected the sweep changes th...

Page 25: ...RANGE screen At the shortest sweep times i e the fewest steps with the widest frequency spans the frequency changes will be quite large at each step if the output is applied to a filter for example th...

Page 26: ...nd from a remote interface With external selected the generator output runs when the signal at the TRIG GATE IN socket is high this input has a TTL level 1 5V threshold The minimum pulse width that ca...

Page 27: ...otary control Tone Gate Source The tone frequency steps changes when the specified trigger signal goes high and continues until the level changes again at which point the output immediately returns to...

Page 28: ...ENCIES screen called by pressing the FSK set up soft key on the MODE screen FSK FREQUENCIES source as gate setup f0 1 000 kHz f1 10 0000 kHz With each frequency selected in turn filled diamond the fre...

Page 29: ...ange 10dB and the full amplitude range is achieved by switching in up to five 10dB attenuation stages Peak modulation cannot exceed the maximum of the range within which the output has been set by cho...

Page 30: ...defaults soft key Note that loading the defaults does not change the set ups stored in memories 1 to 9 or the RS 232 USB interface settings Warnings and Error messages The default setup is for all war...

Page 31: ...etting restore last setup i e the settings at power down are restored at power up or any of the settings stored in non volatile memories 1 to 9 Default values restores the factory default settings see...

Page 32: ...counter to the AUX OUT Frequency meter accuracy will determine the accuracy of the generator s clock setting and should ideally be 1ppm Calibration Procedure The calibration procedure is accessed by...

Page 33: ...sing previous decrements back to the previous step Alternatively pressing exit returns the display to the last CAL screen at which the user can choose to either save new values recall old values or ca...

Page 34: ...hat everything progresses in an orderly way Only the following commands should be used during calibration WARNING Using any other commands while in calibration mode may give unpredictable results and...

Page 35: ...ress for operation on the RS 232 interface is set via the REMOTE menu called using the remote soft key on the UTILITIES screen see System Operations section REMOTE interface RS 232 address 05 baud rat...

Page 36: ...e in the connector at the PC end between pins 1 4 and 6 and between pins 7 and 8 see diagram Pins 7 and 8 of the instrument must not be connected to the PC i e do not use a fully wired 9 way cable Bau...

Page 37: ...hen an instrument is first powered on it will automatically enter the Non Addressable mode In this mode the instrument is not addressable and will not respond to any address commands This allows the i...

Page 38: ...Address for any instrument 14H Talk Address followed by an address not belonging to this instrument 03H Universal Unaddress control code 04H Lock Non Addressable mode control code 18H Universal Device...

Page 39: ...ons in exactly the same way as a standard port Note that it is necessary to set the virtual COM port to the same Baud rate as the instrument being controlled in exactly the same way as with a standard...

Page 40: ...f those which are specified as addressable RS 232 control codes WHITE SPACE is ignored except in command identifiers e g C LS is not equivalent to CLS The high bit of all characters is ignored The com...

Page 41: ...number nrf thus defining the end of the list TONEFREQ nrf1 nrf2 Set tone frequency number nrf1 to nrf2 Hz FSKFREQ0 nrf Sets FSK frequency 0 to nrf Hz FSKFREQ1 nrf Sets FSK frequency 1 to nrf Hz SWPST...

Page 42: ...the instrument parameters to their default values see DEFAULT INSTRUMENT SETTINGS RCL nrf Recalls the instrument set up contained in store number nrf Valid store numbers are 0 9 Recalling store 0 set...

Page 43: ...ed personnel in conjunction with the service manual which may be purchased directly from the Manufacturers or their agents overseas Cleaning If the instrument requires cleaning use a cloth that is onl...

Page 44: ...rted 10 DC Offset level may cause clipping 12 DC only setting will have no effect 13 DC offset changed by amplitude 15 Symmetry has no effect on this wave 16 Manual sweep mode not selected 24 Instrume...

Page 45: ...aults by the remote command RST Main Parameters Std Wave Sine Frequency 10kHz Output 4 0Vpp Output Off DC Offset 0V Zout HiZ Gate Trigger Parameters Source Internal Period 1ms Sweep Parameters Start F...

Page 46: ...Ltd Glebe Road Huntingdon Cambridgeshire PE29 7DR England Telephone 44 0 1480 412451 Fax 44 0 1480 450409 e mail sales tti test com International website www tti test com UK website www tti co uk Book...

Reviews: