1. Product Description
Tiger 300 Design
●
Compatible to KaVo® MULTIflex®
W&H® Roto Quick®
NSK® QD-J®
●
Head size (T: torque, M: mini)
●
Spray hole (single, triple)
●
Press chucking
2797
www.ttbio.com
Tiger 300 Series
High - Speed Handpiece Operation Manual
4. Installation
1. Please follow the specification and properly use
KaVo
®
MULTIflex
®,
W&H
®
Roto
Quick
® and
NSK
®
QD-J
® coupling.
2. Please follow original manufacturers' operation manuals when use the above coupling.
※
KaVo
®
& MULTIflex
®
are the registered trademarks of Kaltenbach & Voigt GmbH CO. ,Germany
。
※
W & H
®
& Roto Quick
®
are the registered trademarks of W&H Dentalwerk Burmoos GmbH S.A. Austria
。
※
NSK
®
& QD-J
®
are the registered trademarks of NAKANISHI INC. Japan
。
1. Information in this operation manual were carefully reviewed. We cannot assume any
liability for any damage arising from improper use of the handpiece or from use of this
manual.
2. Products are subject to change without notice.
3. Indications for Use: Tiger 300 Series High Speed Handpieces are intended for removing
carious parts of tooth, proceeding cavity and crown preparations.
4. Please refer to the bar code on the packing box for manufacturing and expiration date.
10. Disclaimer
CAUTION: U.S. Federal law resticts the use of this device to licensed professionals.
(Rx Only)
3. Safety Notes
1.
Once handpiece is damage (ie deformed caused by hit) or improperly use (ie sterilized
above suggested temperature), be sure not to continue using it.
2.
Before operation, always make sure the product and its accessories meet the specified
requirements (ie drive air pressure. lubrication...etc).
3. Please contact our dealer or us with proof of purchase,
wheneve
r you
hav
e problems such
as abnormal vibration, exceeding temperature or unusual noise...etc.
4.
Avoid
contact between the instrument head and soft tissue (risk of burning due to the
press button heating up).
6. Replacement of Rotor
1. Please contact local dealers and let the technicians replace the rotor.
2. Do not use unsuitable rotors or spare parts, otherwise could damage the handpiece and
result in severe consequence during treatment.
5. Changing Bur
1. Follow the arrowhead direction show in Fig. to take off or install the
bur.
2. Put on protective gloves to reduce the risk of interactive inflection,
especially not to touch the used burs without proper protection.
3. Make sure bur is tightly secured before operation, otherwise risks of
injury to doctors, patients or the third party could occur.
4. Bur must be complied with ISO requirement. Burs not in good
conditions such as being bent and damaged, or are deviated from
ISO standards and in improper length may decrease the lifetime of
the handpiece or even cause severe consequence such as injury to people.
5. Do not press the button while bur is rotating, otherwise it could harm internal parts.
7. Cleaning
1. Put on protective gloves to detach the handpiece from the coupling.
2. Use clean cloth with mild cleaning solution to clean the handpiece
surface carefully, especially handpiece groove, and remove the
residues.
3.
Do not immerse handpiece or cartridges/rotor in disinfectant
solution or water, not to be cleaned by ultrasonic process.
4.
Cleaning solution must be non-erosive and harmless to human. Solution with 80%
ethyl alcohol is recommended. Please also follow recommendation of the cleaning
solution manufacturer for suggested usage.
5. Clean the spray nozzle with the nozzle cleaner needle when it is blocked (see above
figure)
8. Lubrication and Maintenance
Fig. 1
Fig. 2
Fig. 3
1.
Use dental handpiece lubrication oil for maintenance, any unspecified oil not intended
for lubricating handpieces could shorten the lifetime and harm human body.
2. To lubricate 300 series handpiece, put Spray Nozzle into nozzle of the lubrication oil can
(Fig. 1).Connect handpiece with the nozzle (Fig. 2) and inject lubrication oil for 2~3
seconds.
3. Be sure to lubricate the handpieces prior to sterilization and run the handpieces for 20~30
seconds to emit residual oil and dirt.
4. To lubricate handpiece chucks, put Shaft Spray Nozzle into nozzle of the lubrication oil
can and inject lubrication oil into the chuck (Fig. 3) for 1 second. It’s recommended to
lubricate chuck at least once a week and then proceeds step 2.
5. After unpacking and prior to first use, be sure to lubricate with oil and run the handpieces
in low rotation speed for 30 seconds and in full speed for 20~30 seconds to emit residual
oil.
9. Sterilization
TTBIO suggested sterilization in accordance with EN13060
1.
Place handpiece in single-use sterile bag and proceed steam prevacuum sterilization.
Do not open the sterile bag before use in order to avoid contact infection.
2.
Use qualified sterilization equipment specified by the authority/government for
sterilization procedures.
3.
Do not use chemical steam nor immerse handpiece or cartridge/rotor in disinfectant
solution, otherwise could damage the handpiece and cause severe consequences.
4.
Sterilization holding time a minimum of 4 minutes at 134
℃
. (drying time is 15
minutes)
11. Quality and Service
All of our products have passed strict QC procedures before delivery. Please note that
improper use or not using the recommended parts and accessories or modifying our
products without our approval may shorten the lifetime of the product or cause damage.
Please contact us or dealers with proof of purchase for consultation whenever you have
any questions.
2. Specifications
Parameter
※
Air pressure must be higher than water pressure.
※
T: torque head M: mini head
※
Pressure values are measured by multi gauge.
Connection
Rotating speed (rpm)
Bur size-diameter / length (mm)
Drive air pressure (bar)
Suggested drive air pressure (bar)
Water flow rate (ml/min)
Water pressure (bar)
Air pressure (bar)
Single
300K
T/300K
M
300K
3T/300K
3M
KaVo®
MULTIflex®
T
:
300,000
~
360,000
M
:
3
6
0,000
~440,000
Ø1.59 ~ Ø1.60 / T(19-25) ; M(16-21)
2.7~3.3
3
>
50
0.8~2.0
1.0~3.0
W&H®
Roto Quick®
300W
T/300W
M
300W
3T/300W
3M
NSK®
QD-J®
300N
T/300N
M
300N
3T/300N
3M
Triple
Air compressor and other devices used by the handpiece must be in accordance with ISO
requirement. (MUST set air pressure higher than water pressure before operation.)
1. Using contaminated water and compressed air containing moisture, impurity or oil to
drive air could shorten the lifetime of the cartridge/rotor, and block the spray nozzle.
2. If NOT follow the required configuration, the risks of damaging the devices and injury to
people could occur.
3. Handpiece is reusable. After unpacking from package or before operating, please
proceed the sterilization from step 7 to 9 by following the procedures: cleaning>lubrica
-
tion>sterilization.