TTBIO Tiger 300 Series Quick Start Manual Download Page 1

1. Product Description

Tiger 300 Design

 

Compatible to KaVo® MULTIflex®

W&H® Roto Quick®

NSK® QD-J®

 

Head size (T: torque, M: mini)

 

Spray hole (single, triple)

 

Press chucking

2797

www.ttbio.com

Tiger 300 Series

High - Speed Handpiece Operation Manual

4. Installation

1. Please follow the specification and properly use 

KaVo

® 

MULTIflex

®, 

W&H

® 

Roto

Quick

® and 

NSK

® 

QD-J

® coupling.

2. Please follow original manufacturers' operation manuals when use the above coupling.

KaVo

®

 & MULTIflex

®

 are the registered trademarks of Kaltenbach & Voigt GmbH CO. ,Germany

 

W & H

®

 & Roto Quick

®

 are the registered trademarks of W&H Dentalwerk Burmoos GmbH S.A. Austria

 

NSK

®

 & QD-J

®

 are the registered trademarks of NAKANISHI INC. Japan

1. Information in this operation manual were carefully reviewed. We cannot assume any

liability for any damage arising from improper use of the handpiece or from use of this
manual.

2. Products are subject to change without notice.
3. Indications for Use: Tiger 300 Series High Speed Handpieces are intended for removing

carious parts of tooth, proceeding cavity and crown preparations.

4. Please refer to the bar code on the packing box for manufacturing and expiration date.

10. Disclaimer

CAUTION: U.S. Federal law  resticts the use of this device to licensed professionals.

 (Rx Only)

3. Safety Notes

1.

Once handpiece is damage (ie deformed caused by hit) or improperly use (ie sterilized

above suggested temperature), be sure not to continue using it.

2.

Before operation, always make sure the product and its accessories meet the specified

requirements (ie drive air pressure. lubrication...etc).

3. Please contact our dealer or us with proof of purchase, 

wheneve

r you 

hav

e problems such

as abnormal vibration, exceeding temperature or unusual noise...etc.

4.

Avoid

 contact between the instrument head and soft tissue (risk of burning due to the

press button heating up).

6. Replacement of Rotor

1. Please contact local dealers and let the technicians replace the rotor.

2. Do not use unsuitable rotors or spare parts, otherwise could damage the handpiece and

result in severe consequence during treatment.

5. Changing Bur

1. Follow the arrowhead direction show in Fig.  to take off or install the

bur.

2. Put on protective gloves to reduce the risk of interactive inflection,

especially not to touch the used burs without proper protection.

3. Make sure bur is tightly secured before operation, otherwise risks of

injury to doctors, patients or the third party could occur.

4. Bur must be complied with ISO requirement. Burs not in good

conditions such as being bent and damaged, or are deviated from

ISO standards and in improper length may decrease the lifetime of

the handpiece or even cause severe consequence such as injury to people.

5. Do not press the button while bur is rotating, otherwise it could harm internal parts.

7. Cleaning

1. Put on protective gloves to detach the handpiece from the coupling.
2. Use clean cloth with mild cleaning solution to clean the handpiece

surface carefully, especially handpiece groove, and remove the

residues.

3.

Do not immerse handpiece or cartridges/rotor in disinfectant

solution or water, not to be cleaned by ultrasonic process.

4.

Cleaning solution must be non-erosive and harmless to human. Solution with 80%

ethyl alcohol is recommended. Please also follow recommendation of the cleaning
solution manufacturer for suggested usage.

5. Clean the spray nozzle with the nozzle cleaner needle when it is blocked (see above

figure)

8. Lubrication and Maintenance

Fig. 1

Fig. 2

Fig. 3

1.

Use dental handpiece lubrication oil for maintenance, any unspecified oil not intended

for lubricating handpieces could shorten the lifetime and harm human body.

2. To lubricate 300 series handpiece, put Spray Nozzle  into nozzle of the lubrication oil can

(Fig.  1).Connect  handpiece  with  the  nozzle  (Fig.  2)  and  inject  lubrication  oil  for  2~3

seconds.

3. Be sure to lubricate the handpieces prior to sterilization and run the handpieces for 20~30

seconds to emit residual oil and dirt.

4. To lubricate handpiece chucks, put Shaft Spray Nozzle into nozzle of the lubrication oil

can and inject lubrication oil into the chuck (Fig. 3) for 1 second. It’s recommended to
lubricate chuck at least once a week and then proceeds step 2.

5. After unpacking and prior to first use, be sure to lubricate with oil and run the handpieces

in low rotation speed for 30 seconds and in full speed for 20~30 seconds to emit residual

oil.

9. Sterilization

TTBIO suggested sterilization in accordance with EN13060
1.

Place handpiece in single-use sterile bag and proceed steam prevacuum sterilization.

Do not open the sterile bag before use in order to avoid contact infection.

2.

Use qualified sterilization equipment specified by the authority/government for

sterilization procedures.

3.

Do not use chemical steam nor immerse handpiece or cartridge/rotor in disinfectant

solution, otherwise could damage the handpiece and cause severe consequences.

4.

Sterilization holding time a minimum of 4 minutes at 134

. (drying time is 15

minutes)

11. Quality and Service

All of our products have passed strict QC procedures before delivery. Please note that 

improper use or not using the recommended parts and accessories or modifying our 

products without our approval may shorten the lifetime of the product or cause damage. 
Please contact us or dealers with proof of purchase for consultation whenever you have 

any questions.

2. Specifications

Parameter

Air pressure must be higher than water pressure.   

※ 

T: torque head    M: mini head

※ 

Pressure values are measured by multi gauge.

Connection

Rotating speed (rpm)
Bur size-diameter / length (mm)

Drive air pressure (bar)
Suggested drive air pressure (bar)

Water flow rate (ml/min)
Water pressure (bar)
Air pressure (bar)

Single

300K

 

T/300K

 

M

300K

 

3T/300K

 

3M

KaVo®

MULTIflex®

T

:

300,000

~

 

360,000

   M

:

3

6

0,000

~440,000

Ø1.59 ~ Ø1.60 / T(19-25) ; M(16-21)

2.7~3.3

3

50

0.8~2.0

1.0~3.0

W&H® 

Roto Quick®

300W

 

T/300W

 

M

300W

 

3T/300W

 

3M

NSK®

QD-J®

300N

 

T/300N

 

M

300N

 

3T/300N

 

3M

Triple

Air compressor and other devices used by the handpiece must be in accordance with ISO 

requirement. (MUST set air pressure higher than water pressure before operation.)
1. Using contaminated water and compressed air containing moisture, impurity or oil to

drive air could shorten the lifetime of the cartridge/rotor, and block the spray nozzle.

2. If NOT follow the required configuration, the risks of damaging the devices and injury to

people could occur.

3. Handpiece is reusable. After unpacking from package or before operating, please

proceed the sterilization from step 7 to 9 by following the procedures: cleaning>lubrica

-

tion>sterilization.

Dealer

Manufactured by

TTBIO COR

P.

2F.,N

O

.7, 6TH R

O

AD INDUSTRY 

P

ARK 

T

AICHUNG. 

TAIW

AN 40755

TEL:886-4-23595958

E-mail:[email protected]

W

ebsite:www.ttbio.com

Summary of Contents for Tiger 300 Series

Page 1: ...mendation of the cleaning solution manufacturer for suggested usage 5 Clean the spray nozzle with the nozzle cleaner needle when it is blocked see above figure 8 Lubrication and Maintenance Fig 1 Fig 2 Fig 3 1 Use dental handpiece lubrication oil for maintenance any unspecified oil not intended for lubricating handpieces could shorten the lifetime and harm human body 2 To lubricate 300 series hand...

Page 2: ...4 清潔劑必須不含腐蝕性或化學殘留可能 建議使用酒 精含量80 的藥用酒精 5 當噴孔阻塞時 請使用清潔針來進行冷卻水 氣孔暢通工作 以避免因工作溫度過高發生 牙齒傷害之危險 如右圖 1 請戴上保護手套將手機自快速接頭上移除 2 清洗必須使用酸鹼值中性清潔劑 使用清潔的布浸入清 潔劑 仔細擦拭手機的外部和表面 最重要為連接面及 縫隙 務必去除殘留物 3 不可將手機及轉子浸泡在消毒劑或水中 亦不可進行 超音波清洗 七 手機清潔 九 滅菌方法 TTBIO 建議滅菌方法根據EN13060 1 將手機置放於單次使用的消毒袋中進行高溫高壓消毒程式 並全程密封消毒袋 直 到使用前才能拆封 避免中途不必要的接觸感染風險 2 請使用符合您當地政府法規標準之滅菌設備及程式 進行高溫高壓消毒作業 3 滅菌程序請勿使用化學蒸汽或將手機浸泡滅菌消毒液中 否則有發生危險及損壞的 可能性 4 本手機建議滅菌條件為13...