VELOCICALC
®
AIR VELOCITY METER
MODEL 9565 SERIES
OPERATION AND SERVICE MANUAL
P/N 6004851, REVISION F
JUNE 2018
www.
GlobalTestSupply
.com
Find Quality Products Online at:
[email protected]
Page 1: ...VELOCICALC AIR VELOCITY METER MODEL 9565 SERIES OPERATION AND SERVICE MANUAL P N 6004851 REVISION F JUNE 2018 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 2: ...ibrated at the time of its manufacture Instruments returned for calibration are not covered by this warranty g This warranty is VOID if the instrument is opened by anyone other than a factory authoriz...
Page 3: ...GES SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION DISMANTLING OR REINSTALLATION COSTS OR CHARGES No Action regardless of form may be brought against Seller more than 12 months after a cause of acti...
Page 4: ...an Optional Pitot Probe or Airflow straight pitot Probe 9 Thermocouple Ports 10 Connecting the Thermocouples 10 Connecting the Optional Bluetooth Portable Printer Device Models 9565 P and 9565 X only...
Page 5: ...Data Using the Portable Printer Models 9565 P and 9565 X only 36 TrakPro Data Analysis Software 37 LogDat2 Downloading Software 37 CHAPTER 4 MAINTENANCE 39 Recalibration 39 Cases 39 Storage 39 CHAPTER...
Page 6: ...g models Model Description 9565 P Multi Function Ventilation Meter with Differential Pressure Sensor 9565 P NB Multi Function Ventilation Meter with Differential Pressure Sensor No Bluetooth 9565 X Mu...
Page 7: ...communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can...
Page 8: ...s or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment www GlobalTestSupply com Find Quality Products Online at sale...
Page 9: ...be changed Refer to the next section DIP Switch Settings Carbon zinc batteries are not recommended because of the danger of battery acid leakage DIP Switch Settings To access the DIP switch remove th...
Page 10: ...tery compartment is set to NiMH If the DIP switch is set to Alkaline and the AC power adapter is connected then the batteries will be bypassed and the meter will be powered by the AC adapter Be sure t...
Page 11: ...nsor window is fully exposed and the orientation dimple is facing upstream N O T E For temperature and humidity measurements make sure that at least 3 inches 7 5 cm of the probe is in the flow to allo...
Page 12: ...ressures in ductwork Connecting the Static Pressure Probe The Static Pressure probe included with the 9565 P is connected to the port on the 9565 P using the included tubing The Static Pressure probe...
Page 13: ...to the port on the meter N O T E The pitot velocity needs a valid temperature to perform the standard or actual velocity correction This is accomplished in the Actual Std Setup menu If no probe capabl...
Page 14: ...ples The K Alloy thermocouple with mini connector has one terminal wider than the other The wider terminal will be inserted into the bottom of the TC1 or TC2 connector port Thermocouples from an alter...
Page 15: ...h devices Connecting to a Computer Use the Computer Interface USB Cable provided with the Model 9565 to connect the instrument to a computer for downloading stored data C a u t i o n This symbol is us...
Page 16: ...ssed while the AC adapter is attached the instrument instructs you to Unplug the instrument to turn off unit To turn the instrument on again attach the AC adapter or press the ON OFF Key Arrow Keys Pr...
Page 17: ...stant is an averaging period It is used to dampen the display If you are experiencing fluctuating flows a longer time constant will slow down those fluctuations The display will update every second bu...
Page 18: ...nimum of five minutes to let it warm up before zeroing the pressure sensor To zero the pressure reading leave the positive and negative ports open to atmosphere and select the Zero Press The pressure...
Page 19: ...of probe currently connected When set to PRIMARY measurement will be the large font on the display When set to ON measurement will be displayed as a secondary parameter up to 4 can be displayed When s...
Page 20: ...value highlight the Enter Settings option and press the key MENU Zero Press Display Setup Settings Flow Setup VOC Setup Actual Std Setup Data Logging Zero CO Applications Calibration Bluetooth Functio...
Page 21: ...w station manufacturer Up to 5 rectangular ducts 5 round ducts 5 duct areas and 5 Kfactors can be pre programmed for quick use on a jobsite When Flow is set as the Primary measurement in the Display S...
Page 22: ...with this function AM 100 AM 300 AM 600 and AM 1200 If a horn model number is chosen the instrument will return to measuring mode and use a preprogrammed curve to calculate flow rate from velocity wh...
Page 23: ...es temperature plug in probe or thermocouple The Entered Temp range is from 40 to 1832 F 40 to 1000 C MENU Zero Press Display Setup Settings Flow Setup VOC Setup Actual Std Setup Data Logging Zero CO...
Page 24: ...RY and the other to ON Only one can be ON at any given time Log Mode Log Settings The 9565 can be programmed to store measurements to memory in several different logging formats Manual Auto save Cont...
Page 25: ...e mode the user samples are automatically logged to memory at the end of the sampling period To start logging press the key DATA LOGGING Measurements Log Mode Manual Log Settings Choose Test Test 001...
Page 26: ...gging until the key is pressed again DATA LOGGING Measurements Log Mode Auto save Log Settings Choose Test Test 001 Name Test View Data Delete Data Memory SAMPLE TIME 00 05 Min Sec DATA LOGGING Measur...
Page 27: ...starts taking readings by pressing the key The instrument will continue taking samples until the time as set in Test Length has elapsed DATA LOGGING Measurements Log Mode Cont key Log Settings Choose...
Page 28: ...symbol will appear on the display To unlock the keypad press the keys simultaneously The Lock symbol will disappear DATA LOGGING Measurements Log Mode Cont time Log Settings Choose Test Test 001 Name...
Page 29: ...a Analysis Software User s Guide which can be found on the TrakPro software CD which is included with the 9565 TrakPro is a trademark of TSI Incorporated DATA LOGGING Measurements Log Mode Cont time L...
Page 30: ...me using 8 characters maximum Use the arrow keys to move the cursor to a desired location press to accept Repeat until the desired name appears Press SAVE to store custom ID name DATA LOGGING Measurem...
Page 31: ...01 Name Test View Data Delete Data Memory VIEW DATA Choose Test Test 001 View Stats View Samples Print Test CHOOSE TEST Test 001 9 Samples Test 002 7 Samples Test 003 0 Samples Test 004 0 Samples Test...
Page 32: ...se the PRINT command Bluetooth communications must be established between the instrument and the Model 8934 wireless printer or PC set up with Bluetooth communications View Samples TEST 001 Pressure A...
Page 33: ...tooth communications must be established between the instrument and the Model 8934 wireless printer or PC set up with Bluetooth communications TEST 001 Velocity Sample 1 218 ft min Sample 2 280 ft min...
Page 34: ...mory DELETE DATA Delete All Delete Test Delete Sample DELETE DATA Delete All Delete Test Delete Sample DELETE ALL Are you sure YES NO DELETE DATA Delete All Delete Test Delete Sample DELETE TEST Test...
Page 35: ...t the memory available to 100 DELETE DATA Delete All Delete Test Delete Sample DELETE SAMPLE Test 001 14 Samples Test 002 10 Samples Test 003 12 Samples Test 004 8 Samples Test 005 7 Samples Test 006...
Page 36: ...Initiating the Zero CO function will show the sensor CO reading and the time remaining N O T E The Zero CO function should be performed in an area where no combustion is taking place which may affect...
Page 37: ...ng information Draft Rate Application Note TSI 142 Heat Flow Application Note TSI 124 Turbulence Intensity Application Note TSI 141 Percent Outside Air Application Note TSI 138 Leakage Test PANDA Manu...
Page 38: ...s Discover Devices Start the Bluetooth process of finding other devices from the VelociCalc Model 9565 P or 9565 X MENU Zero Press Display Setup Settings Flow Setup VOC Setup Actual Std Setup Data Log...
Page 39: ...00 in order to use 8934 printer AutoConnects Specifies how many times the instrument will attempt to reattach to a paired device after the power is turned on For this option the instrument Discoverabi...
Page 40: ...uter USB port Any USB port can be used LogDat2 Downloading Software The VelociCalc Model 9565 also includes downloading software called LogDat2 LogDat2 software transfers the stored data from the Mode...
Page 41: ...a user s calibration standards The field adjustment is NOT intended as a complete calibration capability For complete multiple point calibration and certification the instrument must be returned to th...
Page 42: ...keypad Keypad locked out Unlock keypad by pressing keys simultaneously Instrument Error message appears Memory is full Download data if desired then DELETE ALL memory Fault in instrument Factory servi...
Page 43: ...ange is a function of actual velocity pressure duct size and K factor Temperature TA Probe Range 14 to 140 F 10 to 60 C Accuracy5 0 5 F 0 3 C Resolution 0 1 F 0 1 C Relative Humidity TA Probe Range 5...
Page 44: ...Range 26 500 samples and 100 test IDs one sample can contain fourteen measurement types Logging Interval Intervals 1 second to 1 hour Time Constant Intervals User selectable Response Time Velocity 200...
Page 45: ...00 ft min 5 m s and are best suited to velocities over 2000 ft min 10 00 m s Range can vary depending on barometric pressure 4 Accuracy is a function of converting pressure to velocity Conversion accu...
Page 46: ...tion 800220 Temperature and humidity probe Indoor Air Quality IAQ Probes Model Description 980 Indoor Air Quality Probe Temperature Humidity CO2 982 Indoor Air Quality Probe Temperature Humidity CO2 a...
Page 47: ...002 Pitot Probe 24 61 cm 5 16 8 mm diameter 634634003 Pitot Probe 36 91 cm 5 16 8 mm diameter 634634005 Pitot Probe 60 152 cm 5 16 8 mm diameter 634634004 Telescoping Pitot Probe 8 to 38 20 cm to 96 c...