Chapter 3

 

Time Constant 

The time constant is an averaging period. It is 
used to dampen the display. If you are 
experiencing fluctuating flows, a longer time 
constant will slow down those fluctuations. 
The display will update every second, but the 
displayed reading will be the average over the 
last time constant period. For example, if the 
time constant is 10 seconds, the display will 
update every second, but the displayed 
reading will be the average from the last 10 
seconds. This is also referred to as a “moving 
average”. 

Log Interval 

The logging interval is a frequency period that 
the instrument will log readings. For example, 
if the logging interval is set to 30 minutes, 
each sample will be the average of the last 30 
minutes. 

Menus 

DISPLAY SETUP

 

Display setup menu is where you will setup the desired parameters to 
be displayed on the running screen. With a parameter highlighted you 
can then use the 

ON

 soft key to have it show up on the running 

screen or select the 

OFF

 soft key to turn off the parameter. Use 

PRIMARY soft key to have a parameter show up on the running 
screen in a larger display. Only one parameter can be selected as a 
primary, and up to 2 secondary parameters can be selected at one 
time. 

PRESSURE ZERO 

To zero the pressure reading, select the Pressure Zero menu. The 
instrument will indicate if the pressure zero was successful. 

SETTINGS

 

Settings menu is where you can set the general settings. These 
include Language, Beeper, Select Units, Time Constant, Contrast, 
Set Time, Set Date, Time Format, Date Format, Number Format, 
Backlight and Auto Off. Use the 

 or 

 soft keys to scroll through the 

settings for each option and use the 

  

key to accept settings. 

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Summary of Contents for AIRFLOW Model PVM620

Page 1: ...MICROMANOMETER ALNOR MODEL AXD620 AIRFLOW INSTRUMENTS MODEL PVM620 OPERATION AND SERVICE MANUAL P N 1980588 REV E FEBRUARY 2016 www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Page 2: ...d by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning g...

Page 3: ...of loss Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY which contains the complete and exclusive limited warranty of Seller This LIMITATION OF WARRANTY AND...

Page 4: ...CHAPTER 3 OPERATION 5 Keypad Functions 5 Common Terms 5 Menus 6 DISPLAY SETUP 6 PRESSURE ZERO 6 SETTINGS 6 FLOW SET UP 7 ACTUAL STANDARD SETUP 7 DATA LOGGING 7 Measurements 7 Log Mode Log Settings 8 D...

Page 5: ...ontainer Check the individual parts against the list of components below If anything is missing or damaged notify TSI immediately 1 Carrying case 2 Instrument 3 Pressure tubing 4 Static pressure tip 5...

Page 6: ...of the danger of battery acid leakage Using the Optional AC Adapter When using the AC adapter the batteries if installed will be bypassed Be sure to provide the correct voltage and frequency which is...

Page 7: ...setting a parameter Press the Menu soft key to select the Menu selections which are Display Setup Pressure Zero Settings Flow Setup Actual Std Set up Data Logging and Calibration Common Terms In this...

Page 8: ...is where you will setup the desired parameters to be displayed on the running screen With a parameter highlighted you can then use the ON soft key to have it show up on the running screen or select t...

Page 9: ...g the or key while on the main measurement screen Make adjustments with the or arrow keys and press to accept or enter the Select Kfactor menu to choose a different pre programmed flow value ACTUAL ST...

Page 10: ...Data Use this to delete all data delete test or delete sample Memory This option displays the memory available Delete All under Delete Data will clear memory and reset the memory available LogDat2TM...

Page 11: ...sing the CALIBRATION menu These field adjustments are intended to make minor changes in calibration to match a user s calibration standards The field adjustment is NOT intended as a complete calibrati...

Page 12: ...10 Chapter 4 Storage Remove the batteries when storing the unit for more than one month to prevent damage due to battery leakage www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Page 13: ...ow or dead batteries Replace batteries or plug in AC adapter Dirty battery contacts Clean the battery contacts Reading fluctuates unstable Fluctuating flow Reposition probe in less turbulent flow or u...

Page 14: ...low Rate Range Actual range is a function of actual velocity pressure duct size and K factor Instrument Temperature Range Operating 40 to 113 F 5 to 45 C Storage 4 to 140 F 20 to 60 C Instrument Opera...

Page 15: ...psi 190 in H2O 360 mmHg 48 kPa 2 Pressure velocity measurements are not recommended below 1 000 ft min 5 m s and are best suited to velocities over 2000 ft min Range can vary depending on barometric...

Reviews: