background image

 

Chapter 2

 

 
Setting-up 

Supplying Power to the Alnor Model AXD620/Airflow 
Instruments PVM620 

The Alnor AXD620/Airflow Instruments PVM620

 

is powered with four 

size AA batteries. 

Installing the Batteries 

Insert four AA batteries as indicated by the diagram located on 
the inside of the battery compartment. The Alnor AXD620/Airflow 
Instruments PVM620 is designed to operate with either alkaline 
or NiMH rechargeable batteries, although it will not recharge 
NiMH batteries. Battery life will be shorter if NiMH batteries are 
used. Carbon-zinc batteries are not recommended because of 
the danger of battery acid leakage. 

Using the Optional AC Adapter 

When using the AC adapter, the batteries (if installed) will be 
bypassed. Be sure to provide the correct voltage and frequency, 
which is marked on the back of the AC adapter. The AC adapter 
will not recharge the batteries. 

Connecting to a Computer 

Use the Computer Interface USB Cable to connect the instrument to 
a computer.  
 
For information on how to download stored data see Chapter 3 

LogDat2™ Downloading Software

 

 

Caution:

 This symbol is used to indicate that the data port 

of the Alnor AXD620/Airflow Instruments PVM620 is 

not

 

intended for connection to a public telecommunications 
network. Connect the USB data port only to another USB 
port. 

 

 

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Summary of Contents for AIRFLOW Model PVM620

Page 1: ...MICROMANOMETER ALNOR MODEL AXD620 AIRFLOW INSTRUMENTS MODEL PVM620 OPERATION AND SERVICE MANUAL P N 1980588 REV E FEBRUARY 2016 www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Page 2: ...d by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning g...

Page 3: ...of loss Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY which contains the complete and exclusive limited warranty of Seller This LIMITATION OF WARRANTY AND...

Page 4: ...CHAPTER 3 OPERATION 5 Keypad Functions 5 Common Terms 5 Menus 6 DISPLAY SETUP 6 PRESSURE ZERO 6 SETTINGS 6 FLOW SET UP 7 ACTUAL STANDARD SETUP 7 DATA LOGGING 7 Measurements 7 Log Mode Log Settings 8 D...

Page 5: ...ontainer Check the individual parts against the list of components below If anything is missing or damaged notify TSI immediately 1 Carrying case 2 Instrument 3 Pressure tubing 4 Static pressure tip 5...

Page 6: ...of the danger of battery acid leakage Using the Optional AC Adapter When using the AC adapter the batteries if installed will be bypassed Be sure to provide the correct voltage and frequency which is...

Page 7: ...setting a parameter Press the Menu soft key to select the Menu selections which are Display Setup Pressure Zero Settings Flow Setup Actual Std Set up Data Logging and Calibration Common Terms In this...

Page 8: ...is where you will setup the desired parameters to be displayed on the running screen With a parameter highlighted you can then use the ON soft key to have it show up on the running screen or select t...

Page 9: ...g the or key while on the main measurement screen Make adjustments with the or arrow keys and press to accept or enter the Select Kfactor menu to choose a different pre programmed flow value ACTUAL ST...

Page 10: ...Data Use this to delete all data delete test or delete sample Memory This option displays the memory available Delete All under Delete Data will clear memory and reset the memory available LogDat2TM...

Page 11: ...sing the CALIBRATION menu These field adjustments are intended to make minor changes in calibration to match a user s calibration standards The field adjustment is NOT intended as a complete calibrati...

Page 12: ...10 Chapter 4 Storage Remove the batteries when storing the unit for more than one month to prevent damage due to battery leakage www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Page 13: ...ow or dead batteries Replace batteries or plug in AC adapter Dirty battery contacts Clean the battery contacts Reading fluctuates unstable Fluctuating flow Reposition probe in less turbulent flow or u...

Page 14: ...low Rate Range Actual range is a function of actual velocity pressure duct size and K factor Instrument Temperature Range Operating 40 to 113 F 5 to 45 C Storage 4 to 140 F 20 to 60 C Instrument Opera...

Page 15: ...psi 190 in H2O 360 mmHg 48 kPa 2 Pressure velocity measurements are not recommended below 1 000 ft min 5 m s and are best suited to velocities over 2000 ft min Range can vary depending on barometric...

Reviews: