6-32
Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
R e t u r n i n g t h e E l e c t r o s t a t i c C l a s s i f i e r f o r
S e r v i c e
The Electrostatic Classifier should be serviced at the factory facility
annually for maintenance of the blower, filters, flowmeter, and other
sensors, and DMAs. An ISO 15900-compliant factory calibration will be
performed at this time. Before returning the classifier to TSI for service,
visit our website at
or call TSI at 1-800-874-2811 (USA)
or 001 (651) 490-2811 for specific return instructions. When you call, have
the following information ready for the Customer Service representative:
Instrument model number
Instrument serial number
A purchase order number (unless under warranty)
A billing address
A shipping address
TSI recommends that you keep the original packaging of the classifier for
use whenever the instrument is shipped, including when it is returned to
TSI for service. Always place protective caps over the inlets/outlets to
prevent debris from entering the instrument. If you no longer have the
original packing material, place the instrument inside a plastic bag to
protect it, and then package with at least 5" (13 cm) of shock
absorbing/packaging material around all sides of the instrument.
C a u t i o n
Do not
ship the Model 3082 Electrostatic Classifier with the lead
shielding, the Neutralizer, or the DMA installed.
Lead shielding
—
Do not
return the classifier with lead shielding
installed, as this may cause damage to the classifier during shipment.
3077/3077A Neutralizer
—Follow safety regulations for correct
shipment of radioactive sources. Consult the appropriate Neutralizer
manual for details.
3088 Neutralizer
—The 3088 Neutralizer can be serviced by TSI, but
should not
be installed during shipment to avoid damage.
DMA
—The 3081A, 3085A, and 3086 DMAs can be serviced by TSI,
but
should not
be installed during shipment to avoid damage.
Chapter 3, "Moving and Shipping the Electrostatic Classifier"