TN-M series crane scale technical manual
- 24 -
10. ERROR CODES
Error Message Description
Solution
Err 4
Zero setting error
Zero setting range exceeded due to
switching on.(4%max)
Make sure platform empty.
--ol--
Over range
Remove the load.
Re calibrate
LO-BAT
Low battery indicator
Please charge the battery
Fail
Calibration Error
Check the test weights & Re calibrate
Err 17
Tare out of range
Minus weight or overload. Remove
the load and restart scale again.
Err 6
A/D value out of range
Make sure platform empty and check
the pan is installed proper. Check the
load cell connectors.