Electrical parts of the table top
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
55
13
Electrical parts of the table top
After replacing motors, always reset the level position and update
the software.
CAUTION
Risk of personal injury due to conductive parts.
Use extreme caution and care when repairing electrical
components. Secure the work area from access by others.
CAUTION
Risk of injury!
Always store tools and removed components securely and make
sure no one can trip or fall over them.
CAUTION
Risk of material damage!
Do not let screws, nuts or other parts fall into the operating table!
Immediately remove any parts that fell in. Parts that have not been
removed can damage other components or cables when the OR
table is adjusted. The operating table may not be released to the
customer if parts fell in and were not removed.
For all screws
≥
M4 without split lock washers, use a medium-
strength screw locking agent.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......