Mechanical parts of the table top
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
37
12
Mechanical parts of the table top
After replacing a gear, always reset the level position of the motor
and update the software.
CAUTION
Risk of injury!
Always store tools and removed components securely and make
sure no one can trip or fall over them.
CAUTION
Risk of material damage!
Do not let screws, nuts or other parts fall into the operating table!
Immediately remove any parts that fell in. Parts that have not been
removed can damage other components or cables when the OR
table is adjusted. The operating table may not be released to the
customer if parts fell in and were not removed.
For all screws
≥
M4 without split lock washers, use a medium-
strength screw locking agent.