Adjustment and calibration tasks
262
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
24
Adjustment and calibration tasks
24.1
Adjusting the spindle
The spindle is telescoping. It consists of two interlocking spindles
that rotate in or out in equal distances. Before each installation, the
telescopic spindle must be adjusted so that both spindles stop at the
same distance before the mechanical end stops during the lifting
function.
The spindle may be adjusted only when it is removed.
1. Remove the 4 screws [2] at the top of the spindle tube [1];
remove the spindle tube [1] and set it aside.
CAUTION
Risk of material damage!
The spindle should never travel under motorized power to either of
the mechanical end stops. During adjustment, the spindle is set such
that it stops before reaching the mechanical end stops during the
lifting function.
If the telescopic spindle is not adjusted or not adjusted correctly, the
inner and outer spindles do not rotate in or out in equal distances.
Serious material damage may result during the lifting function. The
adjustment must be repeated.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......