Safety information for the batteries (#1533137)
24
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
5
Safety information for the batteries (#1533137)
The TruSystem 7000 operating table contains two lithium ion
batteries. Because of their size (number of individual cells plugged
in, amount of lithium contained), they are classified as a hazardous
material. There is a risk of explosion in case of fire. As a result, there
are legal ordinances regarding the handling, deployment, and
installation of this type of battery.
The lithium ion batteries were specially developed by
Trumpf Medical. An electronic protective circuit on a circuit board
in the housing monitors the threshold values of the current / voltage
data. The electrical connection outward is provided through a
7W2 D-Sub socket.
Handling on/in
operating table
The battery is connected using cable W134 (#1538234) for
battery 1 and W172 (#1538235) for battery 2. The battery is
directly connected or disconnected using the corresponding
connection cable in the operating table. There are no special
measures, such as a connection sequence on the power supply. A
Technical parameters
Threshold values
Nominal voltage (3.7 V/cell)
40.7 V
Nominal capacity (2.4 Ah/cell)
4.8 Ah
Working voltage range
37 V…45.1 V
Shutdown undervoltage (2.4 V / cell)
approx. 26 V
Shutdown overvoltage (4.35 V / cell)
approx. 48 V
Shutdown overcurrent
20 A
Self-discharge current with active battery
180
μ
A
Max. discharge current over 2 min.
8 A
Max. continuous discharge current
4.8 A
Charging time with 3 A charge current
Approx. 3 hrs
Storage temperature range
– 40 °C to 80 °C
Max. internal battery temperature during
discharge
60 °C
Max. internal battery temperature during
charge
40 °C
Battery Position
1
on power supply
2
on lift motor
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......