Information plates about the fiber end position
Fig. KR-08142
Position of the fiber end: -4 mm!
If the wrong combination of laser light cable and focusing
optics is used, the destruction of the laser light cable or
the focusing optics may result.
Ø
If there is an information plate "Attention: only for fiber end
position -4" on the plug receptacle of the focusing optics,
the focusing optics may only be operated with a TruFiber
FD34 laser device.
TruFiber FD34 laser devices are characterized by the following
features:
■
"FD34" printed on the nameplate
■
Blue information plate "fiber end position -4 mm" on the LLK
plug
■
This does
not
apply to laser devices TruFiber FD34-50.
Safe combination of laser device, laser
light cable and focusing optics
Characteristics of safe combinations
A combination of laser device, laser light cable and focusing
optics is safe when the following criteria are fulfilled:
Laser device, laser light cable and focusing optics must be
designed for the same wavelength.
Laser light cable and focusing optics must be suitable for the
rated power of the laser device.
12-29-01-A1-CRen
2022-03
Installation
2
‐
17
Numerical aperture (NA)
Position of the fiber end
NOTICE
Wavelength
Rated output
Summary of Contents for KF023
Page 1: ...Operator s manual Programmable focusing optics PFO 33 KF023...
Page 2: ......
Page 6: ...II Good to know 2022 03 12 29 01 A1 CRen...
Page 12: ...0 6 Table of contents 2022 03 12 29 01 A1 CRen...
Page 81: ...Protective glass monitoring active Fig 112272 12 29 01 A1 CRen 2022 03 Installation 2 43...
Page 90: ...2 52 Ambient conditions 2022 03 12 29 01 A1 CRen...
Page 112: ...3 22 Software 2022 03 12 29 01 A1 CRen...
Page 142: ...4 30 Adjusting the observation optics 2022 03 12 29 01 A1 CRen...
Page 170: ...5 28 Cooling water 2022 03 12 29 01 A1 CRen...
Page 176: ...6 6 Informing the service department 2022 03 12 29 01 A1 CRen...