Edit a Partial Special Calibration
If a partial calibration was saved earlier, the gauge will display a
P
beside the option entered before. Select the value to be completed.
If the value already entered is selected, the gauge prompts the
operator to overwrite. To overwrite the partial calibration, press the
〈
YES
〉
key and go to page 6–13,
Create a Special Calibration
. To
use the stored partial calibration, press the
〈
NO
〉
key and select the
other value.
To take measurements, see page 6–13. To enter the true density, see
page 6–14.
Erase Special Calibrations
At the
Special Calibration
menu (page 6–11), press
〈
5
〉
. The gauge
displays the
Erase Special Calibration
menu:
From the
Erase Special Calibration
menu, the operator can erase
either a single special calibration or all special calibrations.
To erase a single special calibration, use the arrow keys to scroll
through the special calibrations names displayed on the first line.
When the gauge displays the special calibration to be erased, press
〈
1
〉
. At the
Are you sure . . .
display, press the
〈
YES
〉
key. The
gauge erases the special calibration and returns to the
Ready
screen.
To erase all special calibrations, press
〈
2
〉
at the
Erase Special
Calibrations
menu. At the
Are you sure . . .
display, press the
〈
YES
〉
key. The gauge erases all the special calibrations and returns
to the
Ready
screen.
SP CAL: Your Special
Calib
1 Erase This Sp Cal
2 Erase All Sp Cal
6
. O
FF
SE
TS
&
C
AL
IB
RA
TIO
NS
Summary of Contents for RoadReader Plus 3450
Page 7: ...Model 3450 vii NOTES...
Page 13: ...xii NOTES...
Page 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Page 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 53: ...3 28 NOTES...
Page 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 181: ...Appendix G 4 NOTES...
Page 187: ...Appendix H 6 NOTES...
Page 199: ...Index 12 NOTES...