6–
Enter Moisture Offset (Keypad Entry)
Use standard laboratory methods (for example, oven drying, and so
on) to determine the true moisture content of a sample taken at the
measurement site (
%M
LAB
).
Take a gauge reading at the site. Record the reading (
%M
GAUGE
).
Take multiple samples and measurements, and calculate the average
moisture of the samples. Use this average value for the offset factor
calibration.
Calculate the offset factor (
k
), using the following formula:
k =
%M
LAB
– %M
GAUGE
100 + %M
GAUGE
×
1000
Access the
Offset
menu by pressing the
〈
OFFSET
〉
key. Access the
Moisture Offset
menu by pressing
〈
2
〉
.
Press
〈
3
〉
Change Offset
, then
〈
3
〉
Keypad Entry.
Follow the gauge
prompts to enter the true moisture and gauge moisture reading. The
gauge displays the moisture offset (
k
) and prompts the operator to
save. To save the offset, press the
〈
YES
〉
key, then select one of the
four memory cell locations.
Moisture Offset
K=xxxx
1-Enable
2-Disable
3-Change Offset
Summary of Contents for RoadReader Plus 3450
Page 7: ...Model 3450 vii NOTES...
Page 13: ...xii NOTES...
Page 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Page 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 53: ...3 28 NOTES...
Page 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 181: ...Appendix G 4 NOTES...
Page 187: ...Appendix H 6 NOTES...
Page 199: ...Index 12 NOTES...