WET DENSITY OFFSET
The wet density offset is the difference between the gauge reading
and the actual density. To access the wet density offset functions,
press
〈
OFFSET
〉
to access the
Offset
menu. Then press
〈
1
〉
to access
the
Wet Density Offset
.
The gauge displays the current wet density offset on the second line.
To enable the offset, press
〈
1
〉
. To disable the offset, press
〈
2
〉
.
To enter a new wet density offset, press
〈
3
〉
. The gauge prompts for
the wet density offset.
To change the offset, select the offset sign (positive or negative).
Enter the wet density offset. Press
〈
ENTER
〉
. The gauge enables the
new wet density offset and returns to the
Ready
screen.
Wet Density Offset:
xxxx kg/m3
1-Enable
2-Disable
3-Change Offset
Wet Density Offset
xxxx kg/m3
Select (+/-)
6
. O
FF
SE
TS
&
C
AL
IB
RA
TIO
NS
Summary of Contents for RoadReader Plus 3450
Page 7: ...Model 3450 vii NOTES...
Page 13: ...xii NOTES...
Page 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Page 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 53: ...3 28 NOTES...
Page 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 181: ...Appendix G 4 NOTES...
Page 187: ...Appendix H 6 NOTES...
Page 199: ...Index 12 NOTES...