![Troxler RoadReader Plus 3450 Manual Of Operation And Instruction Download Page 65](http://html1.mh-extra.com/html/troxler/roadreader-plus-3450/roadreader-plus-3450_manual-of-operation-and-instruction_1172544065.webp)
4–12
ASPHALT MODE
To measure the density of asphalt (and hardened concrete of 4
inches or more) follow the steps below:
✓
Select the
Asphalt
mode (see page 3–23).
✓
If desired, enter or enable the Marshall value, voidless density
value, or Marshall value/voidless density pair (see page 3–19).
✓
Prepare the test site as described on page 4–8.
✓
Place the gauge on the test area.
✓
Lower the source rod to the backscatter position. Release the
trigger. Gently tap on the handle over the source rod to lock the
source rod into position. You should hear a
click
when the
source rod locks into position.
✓
Press the
〈
START
〉
key.
►
In the
Manual
depth mode (see page 3–14), the gauge will
prompt for the source rod depth. Enter the source rod depth
using the number keys. For example, with the source rod in the
backscatter position, press
〈
0
〉
. Then, press
〈
ENTER
〉
.
►
In the
Automatic
depth mode, the gauge software reads the
depth strip on the source rod to determine the source rod depth.
✓
The gauge will display the time remaining while taking a
measurement.
NOTE
Prior to countdown, the gauge may display
Please Wait
for five seconds. This occurs if the baseboard electronics
require warmup prior to starting the measurement.
Summary of Contents for RoadReader Plus 3450
Page 7: ...Model 3450 vii NOTES...
Page 13: ...xii NOTES...
Page 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Page 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 53: ...3 28 NOTES...
Page 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 181: ...Appendix G 4 NOTES...
Page 187: ...Appendix H 6 NOTES...
Page 199: ...Index 12 NOTES...