Model 3450
Appendix A–9
SOURCE ROD INSPECTION
To ensure the integrity of the source rod, Troxler recommends
that a qualified Troxler service person inspect the gauge and the
source rod at least once every five years.
This inspection includes
checking for excessive wear, corrosion, or damage that could affect
the safety of gauge operation.
✓
However, as a precaution, ensure that the tip of the source rod is
intact and undamaged (that is, ensure that the source is not
missing). Use a radiation survey meter to check the radiation
levels at the gauge base surface (without extending the source
rod). A maximum reading of 10-20 mrem/hr is normal, and
indicates the source is present. However, if the maximum
reading is less than 1 mrem/hr or if a survey meter is not
available, perform a visual inspection of the source rod tip as
follows to confirm its integrity:
1.
Extend the source rod just far enough to see the source rod
tip. The tip should appear flat to slightly rounded and
smooth.
2.
Stay at least three feet away from the tip of the unshielded
source rod and complete the inspection as quickly as
possible to minimize exposure (the dose rate at three feet
from the unshielded source is about 2.7 mrem/hr).
If the visual inspection indicates that the source rod tip is
broken off (source is missing):
1.
Immediately contact your Radiation Safety Officer (RSO).
2.
Initiate a search for the source starting at the location where
the gauge was last used.
3.
Report lost or missing radioactive sources to the state or
federal radiation control agency in accordance with
applicable regulatory requirements.
Contact the Troxler Radiation Safety Department for further advice.
A. R
AD
IA
TIO
N T
HE
OR
Y
Summary of Contents for RoadReader Plus 3450
Page 7: ...Model 3450 vii NOTES...
Page 13: ...xii NOTES...
Page 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Page 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 53: ...3 28 NOTES...
Page 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 181: ...Appendix G 4 NOTES...
Page 187: ...Appendix H 6 NOTES...
Page 199: ...Index 12 NOTES...