D-4
Sample analysis must be performed by a
licensed laboratory only.
WARNING!
Ensure that the source rod is in the SAFE position.
L
L
e
e
a
a
k
k
T
T
e
e
s
s
t
t
i
i
n
n
g
g
Routine leak testing of the Model 4590 is not required in the
USA due to the low activity of the source.
However, if you
desire to perform a leak test of the Model 4590, please perform
the following using the Troxler Model 3880 Leak Test Kit (P/N
102868) and accompanying instructions.
1
1
.
.
Write the date, gauge model number, and serial number on
the sample form and label.
2
2
.
.
Ensure that the source rod is in the SAFE position.
3
3
.
.
Turn the gauge on its side and locate the opening where the
source rod extends through the gauge base.
4
4
.
.
Holding the wipe disk with tongs, wipe the area around and
inside the opening where the source rod extends from the
gauge base.
5
5
.
.
Pack the disk, as instructed, in the envelope and mail to
Troxler Electronic Laboratories, Inc. for analysis.
6
6
.
.
Secure the gauge properly.
NOTE
Summary of Contents for EGauge 4590
Page 9: ......
Page 18: ...xvi NOTES...
Page 24: ...1 6 Figure 1 4590 Gauge and Standard Accessories...
Page 26: ...1 8...
Page 70: ......
Page 94: ...6 4...
Page 104: ......
Page 164: ......
Page 168: ......
Page 171: ...Model 4590 F 3 Standard Count Log Gauge Serial Number Date DS BGD Date DS BGD...
Page 174: ......
Page 186: ...Notes NOTES...