![Troxler 3430 Plus Manual Of Operation And Instruction Download Page 129](http://html1.mh-extra.com/html/troxler/3430-plus/3430-plus_manual-of-operation-and-instruction_1172534129.webp)
Model 3430 Plus & 3440 Plus
Appendix A–7
Figure A–2. Effect of Distance on Exposure
Shielding
Shielding is any material used to reduce the radiation exposure rate
from a radioactive source. The gauge has some built-in shielding,
which reduces the exposure rate. When gauges are in storage,
additional shielding may be necessary to keep exposure to personnel
in adjacent areas below the dose limits for members of the public.
The thickness of any material that reduces the incident radiation
intensity by one-half is known as the
half-value layer (HVL)
. The
HVL of a material varies with the type and energy of radiation. The
HVL values of certain common shielding materials are shown
below for gamma and fast neutron radiation.
Material
Cs-137 Gamma
Half-Value Layer
Am-241:Be Neutrons
Half-Value Layer
Concrete
1.9 inches
4.3 inches
Lead
0.3 inches
*
*
Lead does not provide any effective shielding of fast
neutrons.
A. R
AD
IA
TIO
N T
HE
OR
Y
Summary of Contents for 3430 Plus
Page 3: ...Model 3430 Plus 3440 Plus iii ...
Page 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Page 10: ......
Page 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Page 16: ...xiv NOTES ...
Page 30: ...2 6 NOTES ...
Page 64: ...4 NOTES ...
Page 98: ...7 NOTES ...
Page 122: ...9 NOTES ...
Page 142: ...Appendix A 20 NOTES ...
Page 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 190: ...Appendix F 8 NOTES ...
Page 194: ...Appendix G 4 NOTES ...
Page 200: ...Appendix H 6 NOTES ...
Page 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Page 216: ...Index 12 NOTES ...
Page 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...