B-2
MEASUREMENT SPECIFICATIONS
DENSITY AT 2000 KG/M
3
Direct Transmission (150 mm)
.25 min
1 min
4 min
Precision (kg/m
3
)
±
6.8
±
3.40
±
1.70
Composition error (kg/m
3
)
±
20.0
±
20.0
±
20.0
Surface error (kg/m
3
)
–17.0
–17.0
–17.0
(100% Void)
Backscatter (98%, 100 mm)
.25 min
1 min
4 min
Precision (kg/m
3
)
±
16.0
±
8.00
±
4.00
Composition error (kg/m
3
)
±
40.0
±
40.0
±
40.0
Surface error (kg/m
3
)
–75.0
–75.0
–75.0
(100% Void)
MOISTURE AT 240 KG/M
3
.25 min
1 min
4 min
Precision (kg/m
3
)
±
10.3
±
5.1
±
2.6
Precision is defined as
±
one (1) standard deviation in density
readings. This number is calculated by the ratio of the standard
deviation in the counting rate and the slope of the calibration curve
at a given density.
Summary of Contents for 3430-M
Page 9: ......
Page 14: ...xii NOTES...
Page 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Page 22: ...1 8 NOTES...
Page 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Page 48: ...3 20 NOTES...
Page 62: ...4 14 NOTES...
Page 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Page 88: ...A 14 NOTES...
Page 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Page 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Page 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Page 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Page 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 140: ...Index 8 NOTES...