DAILY INSPECTION
The gauge should be inspected daily before use to ensure proper
operation of all safety features as follows:
✓
Push the source rod down into the backscatter position, and then
raise it back to the
SAFE
(shielded) position. The source rod
opening in the bottom of the gauge is equipped with a spring-
loaded tungsten sliding block that shuts when the source rod is
in the
SAFE
position. Turn the gauge over and verify that the
sliding block is completely shut. If any portion of the opening is
uncovered, the sliding block should be cleaned before using,
transporting, or storing the gauge. Refer to the
section of Appendix C for instructions on cleaning
✓
If a radiation survey instrument is available, verify that the
radioactive gamma source is in place by measuring the exposure
rate at the surface of the gauge. The exposure rate should be
approximately 10 – 20 mrem per hour. A reading of about 1
mrem or less indicates either that the survey instrument is not
working properly or that the cesium-137 source may be missing.
Refer to the
section of Appendix C for further
CAUTION
Do not store or transport the gauge unless the
sliding block is completely closed. Increased
radiation levels may violate transportation
regulations and cause excessive personnel radiation
exposure.
G
AUG
E
O
P
E
RAT
IO
N
Summary of Contents for 3430-M
Page 9: ......
Page 14: ...xii NOTES...
Page 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Page 22: ...1 8 NOTES...
Page 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Page 48: ...3 20 NOTES...
Page 62: ...4 14 NOTES...
Page 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Page 88: ...A 14 NOTES...
Page 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Page 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Page 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Page 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Page 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 140: ...Index 8 NOTES...