Model 2701-B
3–7
Enable a Mix Calibration
To enable a displayed mix calibration, press the
corresponding numeric key. The gauge enables the mix
calibration and returns to the
Setup
menu.
New Mix Calibration
To determine a new mix calibration or to manually enter the
slope and intercept values, press
〈
5
〉
on the menu shown on
page 3–6. The gauge displays:
To determine a new mix calibration based on gauge
measurements and values determined by an alternate
method, press
〈
1
〉
to access the
Data Pairs
function. To enter
slope and intercept values manually, press
〈
2
〉
to access the
Slope/Intercept
function. These functions are described in the
following sections.
Data Pairs
To determine a new mix calibration based on gauge
measurements and values determined by an alternate
method, press
〈
1
〉
at the
Entry Method
menu shown above.
The gauge displays:
- Ent r y Met hod-
1. Dat a Pai r s
2. Sl ope/ I nt er cept
I nput Readi ng #: 1
0. 0
pcf
Pr ess <ENTER>
4. U
SIN
G T
HE
G
AU
GE