background image

TRT-BA-BQ30-TC2018-32-001-EN

BQ30

EN

OPERATING MANUAL

PARTICLE MEASURING DEVICE

Summary of Contents for 3510205098

Page 1: ...TRT BA BQ30 TC2018 32 001 EN BQ30 EN OPERATING MANUAL PARTICLE MEASURING DEVICE ...

Page 2: ...rmity via the following link BQ30 https hub trotec com id 42166 Safety Read this manual carefully before starting or using the device Always store the manual in the immediate vicinity of the device or its site of use Warning Read all safety warnings and all instructions Failure to follow the warnings and instructions may result in electric shock fire and or serious injury Save all warnings and ins...

Page 3: ...g from the mains socket and the battery from the device Hold onto the mains plug while pulling the power cable out of the mains socket Warning of explosive substances Do not expose the battery to temperatures above 45 C Do not let the battery come into contact with water or fire Avoid direct sunlight and moisture There is a risk of explosion Warning of laser radiation Laser class 2 P max 1 mW λ 65...

Page 4: ...ss concentration of the contained particles Particles sized 2 5 μm and 10 0 μm are treated equally during this process The detected values for all active particle sizes are simultaneously displayed on a colour display Additionally the air contamination is indicated on a colour scale With respect to the mass concentration of particle size PM2 5 you can define limit values preset in the device which...

Page 5: ...icles sized up to max 10 µm 10 Measurement unit ppm or μg m 11 Indicator scale for particle contamination Colour indicator for the degree of particle contamination For an explanation of the colours please refer to chapter Operation 12 Measured relative humidity value 13 Measured ambient temperature value 14 Air quality assessment Levels good normal poor 15 Measurement status Device performs partic...

Page 6: ...centration measuring range 0 to 2000 μg m Mass concentration resolution 1 μg m Light source of the measuring cell Laser class 1 class 3R laser enclosed in a way to prevent tampering 780 nm 1 5 3 mW classified according to DIN EN 60285 1 and the Technical Regulations of the Occupational Health and Safety Ordinance on Artificial Optical Radiation TROS CO2 concentration CO2 concentration measuring ra...

Page 7: ...necessary The storage temperature is the same as the range given in the Technical data chapter When storing the device for an extended period of time remove the battery batteries Operation Charging the battery The battery is partially charged upon delivery to avoid damage to the battery caused by a deep discharge Warning of electrical voltage Before each use of the charger or power cable check for...

Page 8: ... MEM button 2 to select a submenu Setting the date and time You can set the date and time and select whether to display the time in 24 h or 12 h format To do so please proceed as follows 1 Open the submenu Date Time 2 Press the MEM button 2 to set the desired time format 3 Press the ENTER button 4 once to set date and time 4 You can now set year month day hour and minutes one by one ð Use the butt...

Page 9: ...ollows to have the saved values displayed by the device 1 Press the MEM button for approx 2 s to open the Memory menu ð The last eight measurement results are dynamically displayed in form of a line graph 2 Briefly press the ENTER button 4 to display individual saved measurement results ð Date time and memory slot of the currently selected measurement result will be indicated on the display 1 ð Th...

Page 10: ... www trotec24 com The separate collection of waste electrical and electronic equipment aims to enable the re use recycling and other forms of recovery of waste equipment as well as to prevent negative effects for the environment and human health caused by the disposal of hazardous substances potentially contained in the equipment Li Ion In the European Union batteries and accumulators must not be ...

Page 11: ...Trotec GmbH Co KG Grebbener Str 7 D 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com ...