FLAMMABLE GAS SENSOR/TRANSMITTER
A T E X
M1
GROUP I & II
INTRINSICALLY
SAFE
8 CALIBRATION
continued
8.3
Calibrate ZERO
Apply purge air, at a flow rate of 0.2 to 1 l/min, for 20 seconds
to clear all remnants of gas.
AIR
The CALIBRATE ZERO will NOT function if any level of
gas concentration is detected by the sensing
element – The indicator will NOT flash GREEN.
Press CALIBRATE ZERO until the indicator is flashing GREEN
(about 3 seconds).
Release and the indicator will briefly show RED to denote
ZERO setup.
The Gas Sensing Module will immediately return to the
NORMAL measuring mode.
Switch off the purge air supply.
G
G
R
G
The Gas Sensing Module will automatically return to
the NORMAL measuring mode if NO pushbuttons are
operated within 60 seconds.