9. COMMISSIONING
Once installed and powered for first time use, the Air XS will begin to monitor and sample the
passing environmental particulates. The first-generation Air XS
does not
allow for custom user
parameters or set-up adjustment.
9.1 First power on
Prior to commissioning and first use, the instrument should be inspected for any visible
damage and integrity of the enclosure.
1. Ensure that the electrical connections are correctly installed, as described in
2. Ensure that the door is closed and secured.
3. Apply power to the Air XS.
4. A splash screen will be displayed for several seconds whilst the instrument
auto configures.
5. Once configuration is complete, the Air XS will automatically display on
screen particle data.
6. Data is automatically logged to the internal memory for data download and
offline analysis using the Trolex Breathe application software.
9.2 System shutdown
Note:
Before disconnecting power to the Air XS unit, it is recommended that the system shutdown
procedure is followed correctly. This is to ensure that the onboard operating system and data capture
integrity remains stable.
1. Navigate to the main menu.
2. Scroll to the ‘System Shutdown’ menu selection.
3. Select ‘Confirm’ to begin the system shutdown procedure.
Following this, the device will shut down and the power supply can be
safely disconnected.