If you have chosen “WORK/REST”- rhythm you can set the following
parameters:.
- frequency (Hz)
- pulse width (µS)
- work time (WORK)
- rest time (REST = recovery time)
6. The programming is completed by pressing the “P” button and the program
is stored.
Selectable frequency and pulsewidth settings:
Continuous-Mode: 1 to 150 Hz, 50 to 350 µS
Burst-Mode: 35 to 150 Hz, 50 to 200 µS
Modulated-Mode: 1 to 150 Hz, 50 to 350 µS
Work/Rest-Mode: 1 to 100 Hz, 50 to 350 µS
Safety Adjustments
1. The +buttons become inactive after one minute running time of a program to
avoid an unintentional increase of intensity . This lock is switched off again by
pressing the +button three-times repeatedly. An increasing of intensity is
possible again for another minute.
2. If the electrical circuit should be interrupted during operation (> 6 mA), for
example by losing an electrode or by a broken lead wire, the device stops
stimulating immediately and the display shows the flashing symbol of an
electrode. The effected channel will also flash.