User Manual
9200B 300A AC TRMS Clamp
Page 1: ...User Manual 9200B 300A AC TRMS Clamp ...
Page 2: ... the manual for further information This symbol adjacent to a terminal indicates that under normal use hazardous voltages may be present Double Insulation This meter has been designed to be safe in use but the operator must use caution in its operation The rules listed below should be carefully followed for safe operation SAFETY NOTES Do not exceed the maximum allowable input range of any function...
Page 3: ...ace and fastened securely Meter Description 1 Current Clamp 2 Clamp Trigger 3 Non contact AC Voltage Indicator Sensor 4 Power Button 5 Data HOLD Backlight Button 6 LCD Display 7 Wrist Lanyard Display Description Data Hold Minus sign Negative reading display 0 to 3999 Measurement display digits AUTO Auto Range mode AC Alternating Current Low battery Input Limits Function Maximum Input AC Current 30...
Page 4: ...r or insert into the hot side of the electrical outlet 2 If AC voltage is present the detector light will illuminate NOTE The conductors in electrical cord sets are often twisted For best results rub the probe tip along a length of the cord to assure placing the tip in close proximity to the live conductor NOTE The detector is designed with high sensitivity Static electricity or other sources of e...
Page 5: ...Hold Backlight key for more than 2 seconds Automatic Power OFF In order to conserve battery life the meter will automatically turn off after approximately 15 minutes To turn the meter on again please press ON OFF key again ON OFF BUTTON Power On Pressing the ON OFF button momentarily will turn the meter on Power Off While the meter is on pressing the ON OFF button for more than 1 second turns the ...
Page 6: ...t use a brasives or solvents If the meter is not to be used for 60 days or more remove the battery and store it separately Battery Replacement 1 Remove the Phillips head screw that secures the rear battery door 2 Open the battery compartment 3 Replace the two AAA battery 4 Secure the battery compartment ...
Page 7: ...s 32ºF to 86ºF 0ºC to 30ºC 90 RH 86ºF to 104ºF 30ºC to 40ºC 75 RH 104ºF to 122ºF 40ºC to 50ºC 45 RH Storage conditions 14ºF to 140ºF 30ºC to 60ºC 90 Relative Humidity Altitude Operate at less than 3000 meters Standards IEC61010 Category III 300V Function Range Resolution Accuracy of reading AC Current 50 to 400Hz 40 00A 0 01A 2 8 15 digits 300 0A 0 1A 3 0 15 digits Non Contact AC Voltage 200VAC to...
Page 8: ... from the date of purchase This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as in our sole judgment to injure their stability or reliability or which have been subject to misuse abuse misapplication negligence accident or which have had the serial numbers altered defaced or remo...