41
D01-060en202008 Manual nanoFlu
nanoFlu
//
Annex
Write single register (0x06)
With the “write single register” function, instead of changing the configuration values, a single measurement
trigger is sent, in which a value ≠0 is written.
Designation
Address
Description
Trigger measurement
1
A single measurement is taken.
Report slave ID (0x11)
Provides the sensor designation followed by the serial number followed by the firmware version each as a
null-terminated ASCII character string.
Example:
n
a
n
o
F
l
u 0x00 D
2
0
0 0x00 1
.
0
.
0 0x00
Writing to the configuration registers should be done as seldom as possible and
especially not in every measuring cycle, otherwise the flash memory may be
damaged.
NOTICE
Parametername
RW
603
Char[32]
Name of the parameter
Temperature
R
998
Float
Sensor temperature in [°C] during the last
measurement.
Concentration / scaled
concentration
R
1000/1500
Float