![TriContinent C24000 Product Manual Download Page 31](http://html.mh-extra.com/html/tricontinent/c24000/c24000_product-manual_1162143031.webp)
Initialize pump
C-Series Product Manual
31
Initialize with TCS Pump Commander
1.
Issue the [Z] command.
2.
Observe pump:
–
The plunger arm/holder should move, and the valve will turn. An
“OK” message should appear in the Status: Response window at the
top of the screen. Most of the LEDs at the top of the pump should be
lit. The ERR LED will be off, and the LIFE LED will be blinking slowly.
–
If a “Failed” message is received, check communication settings and
power, or for something blocking the mechanism. The message may
include suggestions for items to check.
–
Once an “OK” message is received, the pump is ready to accept
commands.
Initialize with terminal emulation programs
1.
Issue the [Z] command. An example would be [/1ZR], where /1 rep-
resents the address of the pump (Address Switch setting plus 1).
2.
Observe pump:
–
The plunger arm/holder should move, and the valve will turn.
–
Most of the LEDs at the top of the pump should be lit. The ERR LED
will be off, and the LIFE LED will be blinking slowly.
3.
Issue a [Q] command to confirm initialization, for example [/1Q].
–
If initialization was successful, the response will indicate so.
–
If an error occurred during the initialization, an initialization error
(error 1) will be returned with the response. Check communication
settings and power, or for something blocking the mechanism.
–
Once initialization is successful, the pump is ready to accept com-
mands.