You can use the two slots along the centerline to mount the VMS-1655M backplate directly to a standard single-gang wall box. Use the
backplate as a template to mark the mounting holes and the cable access hole at the center of the backplate.
Mounting location
Flexible tubing cut to
required length
(3 � supplied)
Interior of
fume hood
Exterior of
fume hood
This port must be referenced to
laboratory air for proper orienta�on
and should be located away from
turbulence
To VMS-1655M
Refer to wiring diagram on
page 17 for power supply
termina�on
VMS-1655M sensor interface cable
(10� length, pre-wired at factory)
Figure 9: Alternative vortex sensor mounting location
Figure 10: Display mounting hole pattern