VMS-1655M
LIT-12013297
7
Mount the vortex sensor reference plate on the hood away from
where it might encounter airflow turbulence. Connect the supplied
tubing to the reference port of the sensor and then to the barbed
fitting on the sensor reference plate.
Flexible tubing cut to
required length
(3 � supplied)
This port must be referenced to
laboratory air for proper
orienta�on and should be located
away from turbulence
To VMS-1655M
Interior of
fume hood
Exterior of
fume hood
Refer to wiring diagram on
page 17 for power supply
termina�on
VMS-1655M sensor interface cable
(10� length, pre-wired at factory)
Vortex sensor mounting location
Figure 7: Hood location
Figure 8: Preferred vortex sensor mounting