1.3 Device Control
40
1.3.10. Setting of Digital Parameters and Alarm Thresholds
Setting the values of digital parameters and alarm thresholds in the device is performed in the
same way. To set the alarm thresholds of all displayed digital parameter, click on the displayed
value of the top (or bottom) threshold of the corresponding parameter:
You will see the window for setting the alarm threshold (for example, the window for setting the
top alarm threshold by the Et value of the multigas module):
The lowest value
The highest value
Alarm on/ off
Setting the alarm thresholds for the mean NIBP pressure (MAP) value and
for the ST segment offset for all leads (StDev) is performed in the menu of
general parameters (p. 1.3.8).
There are several ways to set the required threshold value:
Select the required digit of the digital threshold value for correction by clicking on it, it
will be highlighted in a different color. Use buttons to set the required value. In the
same way, set the values of the remaining digits.
Set the lowest possible value by pressing the button with this value displayed.
Set the maximum possible value by pressing the button with this value.
Turn on / off the threshold by pressing the button
“
OFF/ON
”.