4.12 Electric Stretcher Nurse Control Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 36
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
CAUTION:
Use care when removing the individual nurse control switch wires.
Damage to the terminals could occur if care is not used.
4. Remove the nurse control unit (D), and remove the individual nurse control
switch wires (C) from the defective switch.
5. Squeeze the inside grips and pop the switch (E) from the faceplate to
remove any defective nurse control switch (two types).
Replacement
1. Press the new switch (E) into the opening on the faceplate of the nurse
control unit (D) to replace any defective nurse control switch.
2. Connect the individual nurse control switch wires (C) (see figure 4-15 on
page 4-35) to the new switch as shown (see figure 4-16 on page 4-36).
Figure 4-16. Electric Stretcher Nurse Control Switch Wire Diagram
3. To attach the nurse control unit (D) to the frame, secure the unit with the
two screws (B) and the cable clamp (A) (see figure 4-15 on page 4-35).
4. Plug the stretcher into an appropriate power source, and check for proper
operation.
m115a088