4.9 OB/GYN Stretcher Patient Grip Handle
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 27
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
4.9
OB/GYN Stretcher Patient Grip Handle
Tools required:
Ratchet
T30 Torx®
1
head bit
Torque wrench 0-250 in-lb
Drill
3/16" drill bit
Hammer
1/8" pin punch
Removal
1. Complete end tube removal step 1 and step 2 of procedure 4.7).
2. To ensure easy access to the patient grip handle, slightly angle the end tube
(A) toward the foot end of the bed (see figure 4-12 on page 4-27).
3. Remove the damaged patient grip handle (B) from the end tube (A).
Figure 4-12. Patient Grip Handle
1. Torx® is a registered trademark of Textron, Inc.
m115d176
B
A
B
A